Commissioning and parameterization
FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH
99
4.4.4
Changing the display settings with SOPAS ET
To change the factory settings, open the “MCU” device window, enter the Level 1 password
and call up the “Configuration / Display Settings” directory.
Fig. 91
Directory “Configuration / Display Settings“
*Assignment of measured value index to measured value
Section
Selection field
Significance
Common Display
Settings
Display language
Language version shown on the LC display
Display Unit System
Unit system used in displays
Overview Screen
Settings
Bars 1 to 8
Sensor address for the respective measured value bar in the graphic
display
Measured value
Measured value index for the respective measured value bar
Use AO scaling
If active, the measured value bar of the corresponding analog output is
scaled. If not activated, define the limit values separately
Range low
Values for the separate scaling of the measured value bar,
independent of the analog output
Range high
Содержание FLOWSIC200
Страница 1: ...O PERATING INSTRU CTIO NS FLOWSIC200 Flow Velocity Monitor Installation Operation Maintenance...
Страница 30: ...30 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Product description...
Страница 100: ...100 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Commissioning and parameterization...
Страница 104: ...104 FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH Maintenance...
Страница 121: ...Specification FLOWSIC200 Operating Instructions 8013271 1CJ9 V2 0 2022 01 SICK Engineering GmbH 121...