9
1 Product
Overview
The FC-L1 flow calibrator is able to measure flowrates between 5 and 40 L/min.
The compact and easily portable design makes it easy to carry the unit to measurement
sites for calibrating the flowrate of low volume pump.
The FC-L1 can be used for SIBATA LV-40B, LV-40BR, LV-40BW, or LVS-30 low volume
pumps or LV-250 or LV-250R PM2.5 samplers. (Connecting the flow calibrator to a LV-
250 or LV-250R PM2.5 sampler requires a separately sold hose inlet.)
2 Features
•
Four modes are available for displaying flowrate: 25
°
C at 1 atm, 20
°
C at 1 atm, 0
°
C
at 1 atm, or actual volume.
•
Flowrates under the above conditions do not require user calculation, because they
are calculated internally.
•
Measurement flowrate, temperature, and air pressure values are displayed on the
LCD screen.
•
This product can be used for automatic calibration if used in combination with the
Sibata LV-40B, LV-40BR, or LV-40BW low volume pump, or the LV-250 or LV-250R
PM2.5 sampler, and a dedicated communication cable (communication cable with
software for LV-250/LV-40B, sold separately). This product can also be used for
automatic calibration if used in combination with the LVS-30, and a commercially
available USB A-B cable.
•
In comparison with the previous FC-L1, a new accumulated flowrate display function
has been added. As a result, it is possible to confirm the accumulated flowrate value
either manually or when the configured time has ended.