5-3 Flowrate Conversion Temperature Setting
The flowrate conversion temperature can be set as follows.
Pressing the [ENTER] key causes the numeric display to flash and enables changing
the setting (default setting is 25
°
C at 1 atm).
°
C
→
Converts flowrate based on 25
°
C at 1 atmosphere
°
C
→
Converts flowrate based on 20
°
C at 1 atmosphere
°
C
→
Converts flowrate based on 0
°
C at 1 atmosphere
→
Actual
flowrate
Select the setting using the [UP/SPAN] and [DOWN/ZERO] keys, then press the
[ENTER] key to confirm the setting. Once confirmed, settings are remembered even if
the power is switched OFF.
5-4 Accumulation Time Setting
The accumulation time setting can be configured.
(No time setting configured)
→
Manual setting (default setting)
:
(Time setting configured)
→
Timer
setting
Press the [ENTER] key, which causes the hours display (hours) to flash. Then use the
[UP] or [DOWN] key to set the time (hours). Once again, press the [ENTER] key, which
causes the minutes display (minutes) to flash. Then use the [UP] or [DOWN] key to
change the setting for minutes.
Press the [ENTER] key to confirm the setting. (To return
to the previous screen without changing the setting, press the [MODE/ESC] key.)
Note:
The accumulation time can be configured between 1 minute and 9 hours 59 minutes.
The setting can also be changed by pressing the [ENTER] key when the power
is switched ON (see page 12).
No accumulation time
setting configured
(Hours setting)
(Minutes setting)