Sib Geofiz Pribor Ltd
Figure 14
After this, select the desired field using LEFT and RIGHT (from left to right — day,
month, year, hours, minutes), set the desired value using UP and DOWN and confirm
changes using ENTER. Pressing ESC completes execution of this menu option without
changing the current time and date.
4.4.6.
Setup menu… - Default settings
Execution of this option resets all Tester settings to their initial state (relevant Section of
this Manual is shown in brackets):
Geophone type — the first one in the read-only library (4.4.1.1);
String number — 00000001 (4.4.2.1);
Series — 1 (4.4.2.2);
Parallel — 1 (4.4.2.3);
Cable Resistance per Unit Length — 100 Ohms/km (4.4.2.4);
Interval — 5 m (4.4.2.5);
Lead-in cable length — 10 m (4.4.2.6);
Presence of shunts — Yes (4.4.2.7);
Shunt resistance — 1000 Ohms (4.4.2.8);
Polarity — Off (4.4.3.1);
Noise — Off (4.4.3.2);
Leakage — Off (4.4.3.3);
Resistance — On (4.4.3.4);
Natural frequency — On (4.4.3.5);
Damping — On (4.4.3.6);
Sensitivity — On (4.4.3.7);
Distortion — On (4.4.3.6);
Impedance — Large signal (4.4.3.9);
Autoincrement of string number — On (4.4.4.1);
Continue test with reject — On (4.4.4.2);
Save test data with reject — On (4.4.4.3);
String temperature — Auto (4.4.4.4);
25