ADM-880C 07/20/09
7
4.0 DISPLAY MESSAGES AND PROMPTS
The display messages and prompts displayed by the meter are listed alphabetically within this section. Display messages
and prompts are shown within square brackets throughout the text in other sections of this Instruction Manual. An AirFoil
probe reading in MANUAL reading mode is shown as [FP ± nnnn].
Manual and automatic reading prompts and measurement displays have a battery symbol ( ) as the third or fourth character
of the display. This battery symbol indicates the level of charge in the batteries. This battery symbol is not generally shown
in the text throughout this manual.
If either the STORE or AUTO STORE mode has been selected, the read prompts and measurement displays have a letter
as the first character to the right of the reading sequence number. This letter represents the memory grouping and is not
generally shown in the text throughout this manual. Where it is shown in the manual, it is represented by the letter “G”.
An AirFoil probe reading in either STORE mode is shown as [nnG ± nnnn]. When the charge in the batteries begins to fail,
the letter representing the memory group will be replaced by the symbol for an empty battery.
If the TREND mode has been selected, the measurement displays have an arrow as the first character to the right of the
units abbreviation. When the charge in the batteries begins to fail, the message [LOCHARGE] will be displayed after every
30 readings displayed.
4.1 READ PROMPTS
The following ten prompts all include the term READ, which is a signal for the operator to press the READ key to trigger
the actual measurement.
English Units
CF READ
This display indicates that the meter has been placed in the air flow measurement mode (cfm). This message will
appear automatically upon power up if the flaps plug of the FlowHood is connected to the meter.
/
F READ
This display indicates that the meter has been placed in the temperature measurement mode (
/
F).
FP READ
This display indicates that the meter has been placed in a velocity measurement mode (fpm).
Hg READ
This display indicates that the meter has been placed in the absolute pressure measurement mode (in Hg) with
reference to a vacuum.
In READ
This display indicates that the meter has been placed in the differential pressure measurement mode (in wc).
Metric Units
Bar READ
This display indicates that the absolute pressure readings will be displayed in bars (1 bar = 100 kPa) with reference
to a vacuum.
/
C READ
Display indicates that the temperature readings will be displayed in degrees Celsius (
/
C).
LS READ
This display indicates that the meter has been placed in the air flow measurement mode and will read in liters per
second (L/s).
MS READ
This display indicates that the velocity readings will be displayed in meters per second (m/s).
Pa READ
This display indicates that differential pressure readings will be displayed in Pascals (Pa).
4.2 MEASUREMENT READOUTS
In the following 12 examples, n indicates a number in the displayed result. If no sign is displayed, the result is a positive
number. A negative sign indicates a negative number. English and metric units are shown for each example.
Содержание AIRDATA ADM-880C
Страница 42: ...ADM 880C 07 20 09 37 FIGURE 6 3 VELGRID ASSEMBLY...
Страница 50: ...ADM 880C 07 20 09 45 FIGURE 10 1 FRAME STORAGE FIGURE 10 2 FLOWHOOD IN CASE...
Страница 51: ...ADM 880C 07 20 09 46 FIGURE 10 3 FLOWHOOD ASSEMBLY...
Страница 53: ...ADM 880C 07 20 09 48 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...