Aoralscan 2 User Manual
SHNING 3D TECH Co.,Ltd Copyright
2019 All Rights Reserved
- 34 -
(10) Click Calibration.If the Calibrator was not closely plugged onto the
scanner, a warning message would be popped up. So, make sure the scanner
be plugged into the calibration firmly before next operation.
(11) Click Start. The status bar shows the percentage of completion.
NOTE
Normally the calibration takes approximately 5 minutes.
(12) The following message appears once the calibration is completed.
Click OK to exit.