10.1.3 When There is a Big Error
10.1.4 When the Pen Recording is Deflected
Flow chart 2
Big error
Scaling set
properly?
Contact our dealer where you
purchased the instrument, or our
sales representative.
Display or printed
value matches pen
recording?
Calibration
correctly?
Calibrate again (See
pages 99 to 101)
Set properly (See
7.1.1 “Scaling” on
Page 50)
Pen recording
position of
Zero/Span
properly?
䌎䌏
䌙䌅䌓
䌙䌅䌓
䌙䌅䌓
䌙䌅䌓
䌎䌏
䌎䌏
䌎䌏
Calibrate again
(See 7.2.11 on
Page 86)
䋨
Recording error
䋩
Flow chart 3
Pen recording is deflected (0 or 100%)
Input value and
Signs (+/-) are
correctly?
Range and
Scaling set
properly?
Input signal wired
correctly?
Contact our dealer where you
purchased the instrument, or our
sales representative.
Wire correctly
(See 4.2 on Pages 18, 19)
䌎䌏
䌎䌏
䌎䌏
䌙䌅䌓
䌙䌅䌓
䌙䌅䌓
䋨
Sensor burnout, etc
䋩
Apply a correct input
(See 4.2 on Pages 18, 19)
Set them correctly
(See 7.1.1 on Page 50)