9.3 Replacing Consumables
Replace consumables according to the following the list.
No.
Name
Type
Period
Remarks
Quantity
1
Chart paper
H-10100
33 days
When chart feed speed is
20
䋘
/h
1
2 Ribbon
cassette HPSR001H0002C 6
months
When chart feed speed is
20
䋘
/h
1
3
Cartridge pen
WPSR196A000001A 2 months 1 pen, about 2km on beeline
1
4
Cartridge pen
WPSR196A000002A 2 months 2 pen, about 2km on beeline
1
9.4 Adjusting the Pen Recording Position (Point Calibration)
The following describes how to adjust a pen position on the chart paper.
You are well advised to adjust it annually in order to keep recording accurate.
How to Adjust
It is unnecessary to enter the reference. Leave the recorder as it is and adjust it according to the
instructions in 7.2.11 Point calibration. (See Page 86)
9.5 Calibration (Data Calibration)
Calibrate a process variable input anually in order to keep measurement accurate.
Devices required to calibrate
z
Precision voltage(current) generator (Jointly use a digital voltmeter with accuracy of ±0.02 %
or less as required)
z
Precision dial resistor
䌛
Note
䌝
㽲㩷
When recording with a ribbon, a dot printing color phases out. To ensure clear recording,
replace the ribbon cassette as soon as possible.
㽳㩷
Use the ribbon cassette within one year after its purchase. Due to evaporation of ink, a
printing color phases out as time goes on.
㽴㩷
When tracing faint before the changing period, change the cartridge pen ahead of the period.
䌛
Note
䌝
Prior to calibration, warm up for 30 minutes or more after power-on.