![Shining3D Aoralscan 2 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/shining3d/aoralscan-2/aoralscan-2_user-manual_1258144015.webp)
15
Intraoral Scanner User Manual
Interference immunity
The
Aoralscan 2
is intended for use in the electromagnetic environment specified
below. The customer or user of the
Aoralscan 2
should assure that it is used in
such an environment.
Table 2-4 Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601
test levels
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, a relative
humidity of at least 30% is
recommended.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV
differential
mode
±2 kV
common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Содержание Aoralscan 2
Страница 42: ...42 Intraoral Scanner User Manual View lower jaw View the occlusal effect ...
Страница 45: ...45 Intraoral Scanner User Manual ...
Страница 65: ...65 Intraoral Scanner User Manual Figure 1 Scanning process to support edge sweep optimization ...
Страница 70: ...70 Intraoral Scanner User Manual Figure 1 Selecting the implant area ...
Страница 74: ...74 Intraoral Scanner User Manual ...
Страница 75: ...75 Intraoral Scanner User Manual Scan a three piece full jaw example ...
Страница 83: ...83 Intraoral Scanner User Manual undercut area ...
Страница 85: ...85 Intraoral Scanner User Manual Figure 1 Two jaw tile effect ...
Страница 87: ...87 Intraoral Scanner User Manual Dental Manager Pro for details ...