CONTROL PANEL OVERVIEW (CONTINUED) 

Main Control Panel Continued

 

 

Automatic Start / Reset Switch  

The Automatic Start / Reset switch initiates or aborts the pass box auto cycle used to purge aerobic 
atmosphere from the pass box. Pressing then releasing on the start side (left) initiates the auto 
cycle. Pressing and holding on the Reset side (right) for three seconds aborts the cycle, then 
restores the evacuated pass box to the workspace chamber pressure so that the pass box doors 
can be easily opened. 

 

Pass Box Gas Light 

This light illuminates when anaerobic mixed gas is being injected into the pass box during the auto 
cycle. The pass box Gas light 

does not illuminate

 during manual gas injections. 

 

VAC Light 

The clear VAC pilot light indicates the when the pass box is atmosphere is being evacuated during 
auto cycles. The Vac Light 

does not turn on

 when manually cycling the pass box. 

 

Manual Switch 

The Manual Gas / OFF / VAC switch allows the user to manually cycle the pass box. Pressing the 
switch to VAC (the right) vacuums down the pass box, lowering atmospheric pressure inside the 
box. 

Do not vacuum the pass box down to lower than -20inHg of pressure.

 Doing so risks 

boiling sample media and may damage samples or equipment in the pass box. 

Depressing the switch to GAS (the left) injects AMG drawn from the GAS 1 IN port into the pass 
box. AMG injections increase atmospheric pressure in the pass box. Three iterations of evacuating 
the pass box to -18inHg followed by injections of anaerobic mixed gas back to 0inHg cycles the 
pass box. A slight overpressure of 0.5inHg after the final iteration can help make the pass box 
doors easier to open. 

Do not exceed 05.inHg of positive pressure.

 

Содержание BACTRON300

Страница 1: ...BACTRON 110 120 Volts Installation and Operation Manual BACTRON300 BACTRON600 BACTRON900 Previously Designated BACTRONII BACTRONIV BACTRONIV 900...

Страница 2: ...2 P a g e This page left blank...

Страница 3: ...professional industrial or educational use where the preparation or testing of materials is done at an ambient air pressure range of 22 14 31 3 inHg 75 106 kPa and no flammable volatile or combustible...

Страница 4: ...onnect the Foot Pedal Controls 21 Fill the Manometer 22 Installation Cleaning and Disinfection 22 Place Anaerobic Monitoring Strips 23 Install Shelf Spacers 23 Prepare the Incubators 23 Install the Ar...

Страница 5: ...Usage 63 Sleeve Maintenance and Usage 63 Replacing Sleeve Assembly Components 64 Cleaning and Disinfecting 66 Maintaining the Acrylic Glass Panels 68 Vacuum Pump Maintenance 68 Electrical Components 6...

Страница 6: ...s This can result in injury or death Your BACTRON and its recommended accessories are designed and tested to meet strict safety requirements The workstation is built to connect to a wall power source...

Страница 7: ...ance CONTACTING ASSISTANCE If you are unable to resolve a technical issue with the BACTRON please contact Sheldon Technical Support Phone hours for Sheldon Technical Support are 6am 4 30pm Pacific Coa...

Страница 8: ...rt any damage to the carrier service that delivered the BACTRON If the carton is not damaged open the carton and remove the contents Carefully check all packaging before discarding Save the shipping c...

Страница 9: ...Manometer Side Storage Incubator Pass Box Workspace Chamber Workspace Chamber Incubator Catalyst Cartridge Gas and Vacuum Ports Workspace Incubator Control Panel Side Storage Incubator Pass Box Right...

Страница 10: ...ed on back RECEIVING YOUR BACTRON CONTINUED Workspace Chamber Incubator Catalyst Cartridge Gas and Vacuum Ports Pass Box Right Arm Port Control Panel Manometer Left Arm Port Power Outlet Vacuum Pump P...

Страница 11: ...tion Figure 6 Pass Box Inner Door and Sliding Shelf The pass box style airlock is provided with an integral sliding shelf to facilitate material transfers Small items may be introduced into the chambe...

Страница 12: ...can be introduced into the incubator leaving the device outside Readings taken outside the incubator avoid incubator door openings during verifications and calibrations and eliminate subsequent waits...

Страница 13: ...13 P a g e RECEIVING YOUR BACTRON CONTINUED This page left blank...

Страница 14: ...resent page 15 Check that a suitable electrical outlet and power is present page 16 Install the BACTRON in a suitable location Review lifting and handling instructions page 17 Make sure the BACTRON is...

Страница 15: ...onnections Please page 74 for unit dimensions Caster mounted stands for BACTRONs are available for purchase see page 82 ENVIRONMENTAL DISRUPTION SOURCES When selecting a location to install your BACTR...

Страница 16: ...or these units are 15 amps The wall power source must conform to all national and local electrical codes The unit must be positioned so that all end users can quickly unplug the BACTRON in the event o...

Страница 17: ...ng and damage Note To prevent damage when moving the BACTRON turn each of the four leveling feet completely clockwise LEVELING The BACTRON must be level and stable for safe operation Each BACTRON ship...

Страница 18: ...le Gas Configuration Options AMG is also used to cycle the pass box and arm ports To reduce anaerobic mixed gas consumption during pass box auto cycle purges the BACTRON can be connected to a second c...

Страница 19: ...the filtered gas tubing kit to the barb adapter of the T fitting Attach the black tubing of the kit to the AMG regulator Note Failure to connect the T fitting to both the GAS 1 and 2 ports will interf...

Страница 20: ...re to connect the vacuum pump included with the BACTRON 1 Place the vacuum pump in a well ventilated location adjacent to the BACTRON a The pump must be ventilated to prevent overheating An overheated...

Страница 21: ...bly b Connect the branching lines to the two brass barb adapter fittings on the inner sides of the arm port door assemblies Figure 15 Foot Pedal to Arm Port Assemblies Connect the foot pedal supply li...

Страница 22: ...he factory procedure may not meet the standards of your institutional protocols Cleaning and disinfecting the workstation during installation reduces the chance of microbiological contamination Please...

Страница 23: ...e anaerobes or if the incubator is being run at high temperatures PREPARE THE INCUBATORS Open all incubator doors to create a 0 5 inch 1cm gap The incubator doors must be placed slightly ajar prior to...

Страница 24: ...ion 2 Install each door bottom first inserting the tabs into the slots on the arm ports 3 Tilt the door cover up so that it sits securely in the arm port 4 Turn the locking bar to the horizontal posit...

Страница 25: ...25 P a g e INSTALLATION CONTINUED This page left blank...

Страница 26: ...ntaires Indicates Temperature Rep re temp rature Indicates the Over Temperature Limit system Indique le syst me de d passement de temperature Indicates AC Power Rep re le courant alternatif Indicates...

Страница 27: ...ent in the pass box Indique un environnement ana robie dans le sas Indicates an aerobic environment in the pass box Indique un environnement a robie dans le sas Indicates a injection of gas Indique un...

Страница 28: ...28 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank...

Страница 29: ...Panel Incubator Controls BACTRON300 The Incubator controls on the main panel control the operations of the workspace incubator BACTRON600 The Incubator controls on the main panel control the operatio...

Страница 30: ...heating of the chamber in the event of a failure of the main digital controller For more details please see the Over Temperature Limit System description in the Theory of Operations page 35 The red Ov...

Страница 31: ...ates the when the pass box is atmosphere is being evacuated during auto cycles The Vac Light does not turn on when manually cycling the pass box Manual Switch The Manual Gas OFF VAC switch allows the...

Страница 32: ...in its temperature calibration mode and then enter a display value correction The indicator light labeled Heating Activated illuminates whenever the temperature control system is heating the incubati...

Страница 33: ...ours of usage Additionally the BACTRON is programmed to maintain a mild overpressure in the chamber to prevent the infiltration of external atmosphere including the diffusion of molecular oxygen throu...

Страница 34: ...ugh the sleeve assemblies Incubators BACTRON300 and BACTRON900 are each provided with a cabinet style incubator in the workspace chamber The BACTRON600 and BACTRON900 come with two rotating shelves La...

Страница 35: ...re Gauge and Check Valve The water filled manometer in the workspace chamber serves as a visual pressure gauge and dynamic venting check valve during instances of excess overpressures The manometer is...

Страница 36: ...he back of the BACTRON 2 Insert the power cord plug into the room power source outlet 3 Power the BACTRON by placing the main control panel power switch in the On I position a The power switch and all...

Страница 37: ...as configuration is acceptable An AMG source with a 5 hydrogen concentration must be attached to the GAS 1 IN port 2 Note the reading on the AMG gas regulator a The BACTRON300 requires approximately 2...

Страница 38: ...s will not successfully purge workstation 10 Open and remove the right arm port door 11 Unwrap and install a fresh or reactivated catalyst cartridge on the left wall inside the workspace chamber a The...

Страница 39: ...the arm port knob until the interval decreases b If gas continually pulses into the chamber tighten the right arm port door knob until an injection interval of 1 to 2 seconds is established Continual...

Страница 40: ...on for 2 hours whichever happens first tighten and secure the right arm port door Then carry out the Installing the Sleeve Assemblies Chamber Entry and Verifying Anaerobic Atmosphere procedures on the...

Страница 41: ...ve over the lip of the arm port door a Place the ring on the large end of the sleeve inside the groove on the arm port b Make sure none of the sleeve material is trapped or pinched between the ring an...

Страница 42: ...sleeves will deflate simultaneously c Continue deflating until the definition of your fingernails is visible through the sleeves 4 Flush the sleeves with AMG mixture by depressing the Gas foot pedal l...

Страница 43: ...f oxygen The strips turn pink in the presence of oxygen The more color on the strip the more oxygen is present The strips do not lose color after having been exposed to O2 A strip hold its pink colori...

Страница 44: ...te a controlled leak that aerobic atmosphere can be expelled through See the Establish an Anaerobic Environment procedure steps on page 39 b Allow an additional 50 psi or 30 minutes worth of AMG to fl...

Страница 45: ...ranched supply line is securely connected to the barb adaptors on both arm ports 7 Check that the foot pedal assembly GAS and VAC lines are connected to the GAS OUT and VAC OUT ports on the BACTRON 8...

Страница 46: ...r arm port door 4 Withdraw arms from the sleeves one at a time Door Seal Check Carry out the following steps to make sure the arm port doors are sealed 1 Hold the sleeves away from the BACTRON allowin...

Страница 47: ...the letters SP to indicate the Set Point is about to be shown b The temperature display will then show the flashing adjustable temperature set point Note The display will automatically exit the adjust...

Страница 48: ...ther door Place the sensor probe head as close as possible to the geometric center of the incubator A thermocouple sensor probe sleeve may be taped to the shelving using non stick non marking tape as...

Страница 49: ...falls within the acceptable range of your protocol the display is accurately showing the chamber air temperature The Temperature Verification procedure is now complete b See step 2 if a difference fa...

Страница 50: ...es in the workspace chamber are generating heat If you can find no sources of external or internal temperature fluctuations contact Tech Support or your distributor for assistance End of Procedure Set...

Страница 51: ...outer pass box door while the inner pass box door is opened 4 Always close an open pass box door after offloading or loading items a This safeguards against accidental exposure to oxygen from an inat...

Страница 52: ...o the left GAS to inject AMG into the pass box a Hold the switch in this position until the pass box pressure display reads approximately 4inHg b This completes one iteration 4 Repeat steps 3 and 4 tw...

Страница 53: ...the containers Syringes should be loosened and sealed containers partly opened if permitted by your laboratory or production protocol The pass box sliding shelf can hold and transport up to 252 plates...

Страница 54: ...ladium coated pellets inside the catalyst cartridge should grow warm in the presence of oxygen and hydrogen indicating that the cartridge is ready for use 2 If the cartridge does not heat up when expo...

Страница 55: ...rst beaker of carbon media with the second 3 On day 3 reactivate the first beaker of carbon media a Heat at a minimum of 160 C for at least two 2 hours Place the reactivated scrubber in the chamber b...

Страница 56: ...anufacturing recommends the use of distilled water in the resistance range of 50K Ohm cm to 1M Ohm cm or a conductivity range of 20 0 uS cm to 1 0 uS cm for cleaning and humidifying applications HUMID...

Страница 57: ...57 P a g e OPERATION CONTINUED This page left blank...

Страница 58: ...h Record Incubator Temperature Catalyst Changed Condensate Receptacle Drained AMG Cylinder Pressure Reading Monitoring Strip Changed Date 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24...

Страница 59: ...level in the manometer While the chamber is operating under its normal overpressure of 0 5 inches the water level should be even with the lower of the two reference rings 10 Clean and disinfect the w...

Страница 60: ...ng AMG into the workspace chamber to main overpressure Injections are accompanied by a clicking noise that is the chamber gas solenoid opening then closing Verify Chamber Overpressure Verify that the...

Страница 61: ...the acrylic glass panels arm port doors the outer pass box door back panel as well as the side incubator in the BACTRON600 and 900 The manometer exhaust port on the back of the BACTRON will register...

Страница 62: ...rittleness dryness or cracks are present b Clean the gaskets with warm water and soap if sticky or dirty Dry and seat securely on the pass box mounting frames 6 Close and latch both pass box doors a V...

Страница 63: ...needs to be adjusted Adjusting the chamber pressure switch is a service level procedure Locating leaks See the Locating Leaks entry on page 61 for instructions on using a hydrogen leak detector to pi...

Страница 64: ...uff ring 4 Inspect all components for brittleness dryness or cracks Replace as needed Reassemble the sleeve assembly 1 Pull the sleeve cuff on over the cuff ring 2 Roll an O ring onto the cuff ring ov...

Страница 65: ...65 P a g e USER MAINTENANCE CONTINUED This page left blank...

Страница 66: ...the parts of the equipment or with material contained in it Warning Never clean the unit with alcohol or flammable cleaners Avertissement Ne jamais nettoyer l appareil l alcool ou avec des nettoyants...

Страница 67: ...lable disinfectants that are non corrosive non abrasive and suitable for use on stainless steel painted surfaces and acrylic glass Contact your local Site Safety Officer for detailed information on th...

Страница 68: ...result in a quick loss of elasticity Disable or redirect laboratory UV lighting away from the BACTRON The BACTRON should not be exposed to direct sunlight Verify that your laboratory or workspace envi...

Страница 69: ...ng the acrylic glass panels or metal bulkheads Managing excessive condensation at humidity levels that overwhelm the BACTRON condensate controller depends on either lowering the humidity level in the...

Страница 70: ...le in either door Place the sensor probe head as close as possible to the geometric center of the incubator A thermocouple sensor probe sleeve may be taped to the shelving using non stick non marking...

Страница 71: ...alls outside of your protocol range advance to step 2 Reference Device 2 A display calibration adjustment must be entered to match the display to the reference device See next step Reference Device 3...

Страница 72: ...point with the corrected display value Cooling to Set Point 6 Allow the BACTRON sit for at last one 1 hour undisturbed to stabilize after the incubator has achieved the corrected temperature set point...

Страница 73: ...l repeat steps 3 7 up to two more times a Three calibration attempts may be required to successfully calibrate units that are more than 2 C out of calibration Reference Device 9 If the temperature rea...

Страница 74: ...units produced in the series We reserve the right to alter technical specifications at all times WEIGHT Model Shipping Weight Unit Weight BACTRON300 610lbs 276kg 328 5lbs 149kg BACTRON600 760lbs 344 7...

Страница 75: ...IONS Model Inches W D H Centimeters W D H All Models 12 x 13 5 x 12 30 5 x 34 3 x 30 5 PASS BOX PLATE CAPACITY Model Plates All Models 252 INCUBATOR DIMENSIONS Workspace Chamber Incubator Model Inches...

Страница 76: ...17 6 498 BACTRON600 19 538 BACTRON900 17 6 498 Workspace Incubator Volume Model Cubic Feet Liters BACTRON300 1 4 39 6 BACTRON600 BACTRON900 1 4 39 6 Side Storage Incubator Volume Model Cubic Feet Lite...

Страница 77: ...odel Range Uniformity Workspace Incubator Uniformity Side Incubator BACTRON300 Ambient 5 C to 70 C 0 5 C 37 C BACTRON600 Ambient 5 C to 70 C 1 C 37 C BACTRON900 Ambient 5 C to 70 C 0 5 37 C 1 C 37 C P...

Страница 78: ...d Inlet Type T 12 5Amp 250V 5x20mm 3300520 Arm Port Door Right 9900698 Fuse Vacuum Pump Outlet Type T 10Amp 250V 5x20mm 3300516 Arm Port Door O Ring 6000509 Gas Regulator Kit Anaerobic Mixed Gas 97405...

Страница 79: ...g straps 9990738M Sleeve Cuffs Latex Size 9 for sleeve assembly 3600502 Sleeve Assembly Size 9 Large 2 cuffs 2 cuff rings 4 0 rings 2 sleeves 2 self gripping straps 9990738L Sleeve Cuffs Nitrile Size...

Страница 80: ...322 4897 extension 3 If you are uncertain that you have the correct Part Number or if you need that specific item please contact Sheldon Technical Support for help at 1 800 322 4897 extension 4 or 503...

Страница 81: ...Number 9490578 For 110 120 volt units Acrylic Glass Cleaner 2oz 59 2ml Novus brand acrylic glass cleaner Part Number 1060503 Acrylic Glass Scratch Remover 2oz 59 2ml Helps remove visible scratches an...

Страница 82: ...N900 Microscope Adaptors Designed for the Lecia S6 Spotting Stereo Microscope BACTRON300 Part Number 9990535 BACTRON600 and BACTRON900 Part Number 9990511 BACTRON300 Stand A rolling stand with cabinet...

Страница 83: ...he appropriate BACTRON microscope adaptor Part Number 9990516 Lukas Fiber Optic Micro Lite Illumination System A fiber optic adjustable brightness halogen light box and guide Provides a stable long la...

Страница 84: ...P a g e ACCESSORIES CONTINUED UV Viewing Lamp A handheld UV lamp for use with BACTRON workstations Parts Number 9490507 Zephiran Benzalkonium Chloride Chamber Cleaner 1 Gallon 0 133 Part Number 10605...

Страница 85: ...85 P a g e ACCESSORIES CONTINUED...

Страница 86: ......

Отзывы: