
19 |
P a g e
INSTALLATION (CONTINUED)
C
ONNECT THE
G
AS
S
UPPLY
Warning
: Never exceed a 5% hydrogen concentration in the unit chamber.
Avertissement:
La concentration d’hydrogène ne doit pas dépasser
5%
dans
la chambre anaérobie
.
Single Gas Configuration
Carry out the following steps to connect a single cylinder
of anaerobic mixed gas.
1. Locate the T-shaped brass fitting included with the BACTRON AMG gas regulator
(normally shipped in the regulator box)
2. Connect the fitting to the
GAS 1
and
GAS 2
IN
ports on the BACTRON gas panel
3. Attach the AMG regulator provided with the BACTRON to the cylinder, set to 15
– 20 psi
4. Attach the clear tubing on the filtered gas tubing kit to the barb adapter of the T-fitting.
Attach the black tubing of the kit to the AMG regulator.
Note:
Failure to connect the T-fitting to
both
the GAS 1 and 2 ports will interfere with pass box auto
cycles.
Dual Gas Configuration
Carry out the following steps to set up the dual gas
configuration.
1. Attach the AMG regulator provided with the
BACTRON to the AMG supply cylinder. Set to 15
–
20 psi
2. Connect an appropriate regulator to the neutral gas
cylinder, set to 15
– 20 psi
3. Connect the AMG regulator tubing to the
BACTRON
GAS 1 IN
port
4. Connect the inert gas cylinder supply tubing to the
BACTRON
GAS 2
IN
port
Do not start a flow of gas to the BACTRON at this time, for either configuration.
Figure 8: T-Fitting
Figure 9: Single Gas Configuration
T-Fitting
Filtered Gas
Tubing Kit
Figure 10: Dual Gas Configuration
Содержание BACTRON300
Страница 2: ...2 P a g e This page left blank...
Страница 13: ...13 P a g e RECEIVING YOUR BACTRON CONTINUED This page left blank...
Страница 25: ...25 P a g e INSTALLATION CONTINUED This page left blank...
Страница 28: ...28 P a g e GRAPHIC SYMBOLS CONTINUED This page left blank...
Страница 57: ...57 P a g e OPERATION CONTINUED This page left blank...
Страница 65: ...65 P a g e USER MAINTENANCE CONTINUED This page left blank...
Страница 85: ...85 P a g e ACCESSORIES CONTINUED...
Страница 86: ......