Page 29 of 36
www.sgm-lektra.eu
8.6.2 - SELF CALIBRATION
Position the cursor on self calibration, press ENTER to confirm.
Manually measure in mm the “ACTUAL HEAD” and insert the data, the unit will
automatically calculate the fluid distance to the “Q=0” point (zero flow distance).
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confirm. LEFT ARROW to exit without changes.
PRIMARY DEVICE
SELF CALIBRATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH
PEAK VALUE
START TOTALIZER
ALARM CONFIGURATION
►
0120 mm
SET ACTUAL HEAD
WARNING - Proper programming of this parameter is essential for correct flow measurement.
Do not proceed without having carefully read the below described instructions
ACTUAL
HEAD
ACTUAL
HEAD
ACTUAL
HEAD
Q=0 DIASTANCE
Q=0 DIASTANCE
Q=0 DIASTANCE
FLOWMETER -
configuration
It is recommended to use the “SELF CALIBRATION” system with the zero flow condition, because in doing so the “ACTUAL HEAD”
manually measurement distance errors are avoided.
ACTUAL
HEAD=0
ACTUAL
HEAD=0
ACTUAL
HEAD=0
Q=0 DIASTANCE
Q=0 DIASTANCE
Q=0 DIASTANCE