4 Performing a correlation measurement
Press Cancel to return to the
z
main menu.
Depending on the selected
settings (see section 7.6.2)
either 16, 32, 64 or 128
aver-
aging processes (i.e., meas
-
urements) are performed.
The display shows the current
-
ly performed measurement.
During the measurement, the
correlation function is continu-
ously updated.
The correlation measurement
z
can be interrupted any time
by pressing the ON/OFF key
shortly.
An interrupted measurement
can always be resumed.
After the
z
measurement the re-
sult is displayed. The adjacent
illustration shows a correlation
example. The method of
dis-
play depends on the setting
selected in the „Filter /
Setup“
menu.
4.3
Evaluating the result
The correlation function uses
peaks to indicate where a leak
may be located. The time-delay
difference is displayed on the x
axis in ms (i.e., milliseconds).
The highest peak is automati
-
cally highlighted.