LISST - 25 User’s Guide
14
DT
Display time
Syntax:
DT or dt
Description:
The current time and date are displayed to the screen.
Example:
input: DT
output: Current Date: 04/06/2000
Current Time: 15:07:57
Cautions:
The time is reset to midnight Jan 1, 1980 at power up. To reset time see ST
command.
HE
Display general help messages and command list
Syntax:
HE or he
Description:
Displays the list of command to the screen.
Example:
input: HE
output: LISST-25X Commands (followed by 30 lines of text)
Cautions:
None
Setup Commands
RS
Reset to standard defaults
Syntax:
RS or rs
Description:
Returns all settings to startup defaults. Does not effect datafile, batch file, or
clock. A warning prompt will ask you if you are sure you want to reset to
defaults. Enter “
Y
” or “
y
” followed by a <
enter
> if you wish to reset. Any
other key(or no key) followed an <
enter
> will not reset. After resetting to
default parameters the current status is automatically displayed using the DS
command.
Example:
input: RS
output: Reset to Default parameters? (y or n): “
Y
” <
enter
>
LISST-25X Current Status
…
Cautions:
The calibration constant is an instrument specific value that is determined at
the factory before delivery of the instrument. The default parameter may not
be the correct one for your instrument. Record the current calibration
constant before executing an
RS
command.