LISST - 25 User’s Guide
13
2.5
LISST-25X Command Details
Display Commands
DB
Display battery voltage
Syntax:
DB or db
Description:
Display main battery voltage to the screen. Units are in volts. If battery
voltage is below the suggested level a warning will be displayed.
Example:
input: DB
results: Battery voltage =8.98 volts
Cautions:
Using the instrument with voltage levels below recommended levels could
result in unreliable data.
DM
Display memory remaining
Syntax:
DM or dm
Description:
The memory of the LISST-25X is non-volatile EEPROM. As data is written
to memory, the amount of available memory can be accessed by the DM
command. The amount of available memory is then displayed. The units
will be in number of samples.
Example:
input: DM
output: Memory available: 28103 samples
See also:
The DS, Display Status command will also show the number of bytes
available in the data file.
DS
Display current status information
Syntax:
DS or ds
Description:
The instrument settings and status are displayed to the screen.
Example:
input: DS
output: Current Status of LISST-25X
LISST-25X Serial #25-001
Maximum Memory: 28103 Samples
Memory used: 0 Samples
Memory Available: 28103 Samples
Samples per Average = 10
Calibration Constant = 17500
Sample Rate = 1
Background File Date/Time: 02/23/99 12:03:23
Battery Voltage: 9.23
Current Date: 3/23/99
Current Time: 23:34:45