LISST - 25 User’s Guide
11
2.4
LISST-25X Command Summary
Display Commands
DB
Display battery voltage
DM
Display memory remaining
DS
Display current status information
DT
Display time
HE
Display general help messages and command list
Setup Commands
RB
Hard Reboot to restore factory defaults, time, and
internal calibrations and constants.
RS
Reset to factory defaults settings
SA x
Set samples per average equal to x
SB x
Set communications baud rate to x ,
where x = 300, 600, 1200, 3400, 4800, 9600, 19200,
or 38400
SC x
Set Calibration Constant equal to x
SF x
Set output format where x = sample code
SM x
Set output display mode to x code
SR x
Set sample rate, where x= seconds between samples
ST mm/dd/yyyy hh:mm:ss
Set time and date, where
mm=month,dd=day,yyyy=year,hh=hour(24
format),mm=minute, ss=seconds, Example: ST
01/05/2001 21:05:03
Acquisition/Action Commands
GR
Grab and store samples as set rate
GS
Grab sample and store
GX
Grab sample and transmit it
XS
transmit last sample
ZS
Acquire and store Zscat
ZZ
Go into deep sleep mode (minimum power
consumption)