8. Error messages (Error code)
Messages
Reason
Dispositions
Error1
Asy (Zero-point) exceeds
upper/lower limitation
1.
Please use new buffers.
2.
Clean the electrode or change
to a new electrode, and make
another calibration.
Error2
Slope exceeds upper/lower
limitation
1.
Please use new buffers.
2.
Clean the electrode or change
to a new electrode, and make
another calibration.
Error3
The readout is unstable
1.
Check whether there is bubble or
air in the glass end of the
electrode
2.
Clean the electrode or change
to a new electrode, and make
another calibration.
Error4
1.
The temperature is over
the range 0~50
˚
C during
calibration.
2.
Buffer cannot be
recognized
1.
Please adjust the standard
solution to the proper
temperature range.
2.
Please check whether there is
bubble or air in the glass end of
the electrode
, or maintain the
electrode or change a new
electrode, and make another
calibration.
Error5
Wrong password
ERROR CODE
Re-enter a password
Error9
Serious error that does
not permit any further
measuring
Please contact Sensorex Technical
Support.