3.3 Description of terminal function:
01
02
03
04
05
06
07
NC:
No connection
08
100~240VAC:
Power supply terminal
09
NC
:
No connection
10
100~240VAC:
Power supply terminal
11
12
13
NC:
No connection
14
REF:
Coaxial shield of pH/ORP electrode signal wire
15
SG:
Solution ground wire.
If no solution ground, there should be a
jumper between this terminal and REF (14). A short circuit bar is
attached when shipped from the factory.
16
T/P:
Connect with one of cable end of temperature probe
17
SG:
Connect with one of cable end of temperature probe
18
19
20
4~20mA +terminal
:
Master measure current output te, for external
recorder or PLC control
21
4~20mA
- terminal
:
Master measure current output terminal -, for external
recorder or PLC control
22
4~20mA + terminal/ D+(B)
:
Temperature current output te, for
external recorder or PLC control (only applicable for
TX2000); or RS-485 output D+(B) (only applicable
for TX2000RS)
23
4~20mA - terminal
/
G
:
Temperature current output terminal -, for external
recorder or PLC control (only applicable for TX2000); or
RS-485 output GND (only applicable for TX2000RS)
24
NC / D-(A):
NC or RS-485 output D-(A)
(only applicable for
TX2000RS)or RS-485 output D-(A) (only
applicable for TX2000RS)
WASH:
Wash relay contact for an
external relay
REL2
: Second alarm control, contact for an
external relay
REL1
: First alarm control, contact for an
external relay
GLASS:
Coaxial inner wire of pH/ORP
electrode signal wire
DC
±
12V
:
Output terminal of direct current voltage ±12V