6.9 Clean
Enter setup of “Clean” function. Select the icon to turn on or turn off the clean function. If you
select “Auto” turning on, then set the timer of the clean function including automatically turning on
time and turning off time, and set the Hysteresis value(Hys.).
Note: When the clean function is turned on, if any value is set to 0, the instrument will automatically turn off
this function. When the clean function is activated under measurement mode, there is a “Clean Running”
message showing on top of the display. The measurement value will default to the last measured value
before cleaning. If entering setting menu or calibration menu during clean status, the instrument will stop
clean status automatically.
Press
or
to adjust the auto
turning on time.
Press to
confirm the
minute part, and
move to adjust
the second part.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or to
select to activate a test of
Clean or not. If “ON”, the
Clean function is in
action, and the “ACT”
indication light will light
up.
Press
or
to select to activate
Clean or not. If
not, the guide
menu goes to setup
of Analog 1.
Press to confirm it.
Press or
to adjust Hysteresis
(Hys.) value
Enter “Analog 1” Setup
on off
on off on
off
off
on on
off
T
on
T
hys
T
off
T
on
T
hys
T
off
T
off
Clean Timer Control
Relay
Contact
Timer
Settings
Measure-
ment
Содержание TX2000
Страница 1: ...1 T X 2 0 0 0 Intelligent pH ORP Transmitter Operation Manual KϵϬϬϭͲ2008CertificateNo C0036132 ...
Страница 9: ......
Страница 10: ......
Страница 25: ...3 ...
Страница 30: ...4 ...
Страница 50: ... KϵϬϬϭͲ2008CertificateNo C0036132 ...