6.7 Relay 1
Enter setup of Relay 1. Select the item to turn on or turn of the relay 1 function. If you select to turn
on the relay 1, then select for using relay 1 as “High set-point” alarm or “Low set-point” alarm. Set
the value of set-point (SP) and Hysteresis (Hys.). The relationship between parameters can refer to
an explanatory diagram of the box (as a high point alarm).
High Point Alarm Control
off
off
on
on
SP
Hys
pH
t (sec)
Press
or
to select to activate
REL1 or not. If
not, the guide
menu goes to setup
of Relay 2.
Press to confirm it.
Press to confirm it.
Enter “Relay 2
” Setup
Press
or
to select use REL1
as Hi point or Lo
point alarm.
Press to confirm it.
Press or
to adjust set-point
(SP) value
Press to confirm it.
Press or
to adjust Hysteresis
(Hys.) value
Press to confirm it.
Press to confirm it.
Press
or to
select to activate a test of
REL1 or not. If “ON”, the
relay 1 is in action, and
the “ACT” indication
light will light up.