MANUAL CALIBRATION
The following instructions pertain to manual calibration of the SENSIT®
GOLD G2. If you are using the automatic Smart-Cal Calibration System,
the procedure is different. See the Smart-Cal sections of this manual or
consult the Smart-Cal instruction manual for details.
CARBON MONOXIDE (CO) CALIBRATION (100PPM CO/AIR)
From the working display, press & hold the TICK/MENU BUTTON (B)
until the top line reads USER MENU. Press & release the SAVE/ZERO
BUTTON (C) until the bottom line reads CAL.
Press & release the TICK/MENU BUTTON (B) once. The bottom line
will read AUTO CAL. Scroll with the SAVE/ZERO BUTTON (C) until CO
100PPM is displayed. Apply 100ppm CO/Air calibration gas and press &
release the TICK/MENU BUTTON (B) to start CO calibration.
When the reading is satisfactory, the display will flash DATA SAVED
indicating that calibration is complete for that sensor. The date for CAL
PAST DUE is automatically reset for that sensor as well.
Scroll with the SAVE/ZERO BUTTON (C) if you need to calibrate another
sensor. When finished, remove and shut off the gas supply. Press &
release the POWER/MUTE BUTTON (A) to return to the working display.