LEAK SEARCH 

(OPTIONAL)

To conduct a LEAK SEARCH (LS):

To  enter  the  LEAK  SEARCH  mode  from  the  work  display,  press  and 

release the TICK/MENU (B) button. Press and release the SAVE/ZERO 

(C) button until LS is displayed on the bottom of the screen. Press and 

release the TICK/MENU (B) button. LEAK SEARCH will be displayed on 

the top of the screen with 0 PPM on the bottom screen.

Attach a drag tube assembly or telescoping survey probe. The instrument 

has a preset alarm of 10ppm (adjustable, contact Sensit for details.) The 

instrument will read in 1ppm increments up to 5000ppm, auto range to 

LEL and then to %v/v. 

To zero the instrument in the LS mode, press and hold the SAVE/ZERO 

(C)  button  until  “Autozero”  is  displayed.  Any  alarm  can  be  muted  by 

pressing and releasing the POWER/MUTE (A) button once. If the alarm 

sound is turned off before an alarm condition is met, the alarm will remain 

off  until  activated  by  pressing  and  releasing  the  POWER/MUTE  (A) 

button. If the alarm sound is muted during an alarm condition and the 

concentration of gas is below the alarm threshold, the alarm will activate if 

the concentration exceeds the alarm threshold again.

To exit the LS mode, press and hold the POWER/MUTE (A) button for 2-3 

seconds to exit to the work display.

Содержание GOLD G2

Страница 1: ...atmospheres disconnect power before servicing INSTRUCTION MANUAL SENSIT GOLD G2 Read and understand instructions before use For use with combustible gases and optionally available oxygen and toxic gas...

Страница 2: ...Technologies with any questions For best accuracy always zero in clean air environments similar in temperature and relative humidity to the environment where the instrument will be used When continuo...

Страница 3: ...xygen greater than 21 Warning ONLY zero instrument in a gas free environment Warning To maintain intrinsic safety service must be performed by factory authorized technicians with approved replacement...

Страница 4: ...t Features 11 13 Sensor Types and Pumps 14 Basic Operation Battery Installation Replacement 15 Operation and Use 16 22 Bar Hole Test 23 24 Leak Search Mode Optional 25 Purge Mode Optional 26 Working D...

Страница 5: ...ration Calibration Overview Definitions 44 Calibration Auto Cal 45 46 Carbon Monoxide CO Calibration 47 Hydrogen Sulfide H2S Calibration 48 Hydrogen Cyanide HCN Calibration 49 Methane Calibration 50 2...

Страница 6: ...Vol O2 CO H2S HCN 911 00000 01 911 00000 02 911 00000 03 911 00000 04 911 00000 05 911 00000 06 911 00000 07 911 00000 08 911 00000 09 911 00000 51 911 00000 52 911 00000 53 911 00000 54 911 00000 55...

Страница 7: ...sion Adapter 873 00016 Hydrocarbon Filter 1 883 00023 Hot Air Probe Assembly 883 00021 32 Heavy Duty Fiberglass Probe 883 00015 Confined Space Probe with Tubing 883 00019 32 Fiberglass Probe 873 00013...

Страница 8: ...advanced low power semiconductor sensor to measure combustible gases in the LEL Lower Explosive Limit range and a thermal conductivity style sensor to measure combustible gases in the percent volume...

Страница 9: ...ogen sulfide H2S alarm is preset at 10ppm The hydrogen cyanide HCN alarm is set to 5ppm The SENSIT GOLD G2 instrument is approved ATEX Zone 1 intrinsically safe Ex ib IIB T3 IP54 and UL CLI Div 1 Grou...

Страница 10: ...100ppm 2ppm or 5 HCN 1ppm 0 30ppm 2ppm or 5 Whichever is greater PRODUCT SPECIFICATIONS Size 11 5 x 3 x 2 3 in 292 x 76 x 59 mm Weight 1 2 lb 544 g Operational Temp 4 to 104 F 20 to 40 C Storage Temp...

Страница 11: ...11 PRODUCT FEATURES LUER PROBE CONNECTOR SENSOR CAP LEL PPM SENSOR INFRARED COMMUNICATIONS GRAPHIC DISPLAY ALARM LEDs OPERATION BUTTONS WATER RESISTANT HOUSING FLEXIBLE GOOSENECK BATTERY COMPARTMENT...

Страница 12: ...th the speaker located in the back of the instrument An infrared communication window is located on the right side to allow the Sensit Gold G2 instrument to download calibration data readings the oper...

Страница 13: ...d LED on the right side will flash during any alarm condition In darkened environments the backlight will illuminate in red during alarm conditions There are 3 operational button pads on the front of...

Страница 14: ...and percent volume Electrochemical Sensors optional All Sensit Gold G2 instruments when equipped with the following optional sensors microprocessor and associated circuitry will measure oxygen levels...

Страница 15: ...oin or flat object and pull the handle away from the top or display area of the instrument NOTE Observe the polarity markings on the inside of the battery holder Instrument will not operate with impro...

Страница 16: ...y check vi Microprocessor check vii Pressure sensor check viii Clock check ix Auto Log Check alert at 50 records remaining before memory is full and overwrites c Display all active sensors d Display S...

Страница 17: ...the instrument determines that a sensor is inoperable an ERROR message will flash for that sensor Then FAIL will show on the display for the corresponding sensor Attempt to manually zero using SAVE Z...

Страница 18: ...green LED will flash until adequate flow is present 7 It may be necessary to manually zero the instrument based on company practices and environmental conditions Always zero the instrument in a clean...

Страница 19: ...t assembly 10 When sampling the appropriate sensors will cause the display to update when a gas is encountered Additionally if a combustible gas is encountered a series of LEDs on the front of the ins...

Страница 20: ...50 LEL Methane to 17 volume Methane LED indicator only above 17 volume Methane PROPANE 50 LEL Propane to 12 volume Propane LED indicator only above 12 volume Propane When equipped with percent volume...

Страница 21: ...ating small leaks press and release the TICK MENU BUTTON B The SELECT TEST TICK is displayed Push the TICK MENU BUTTON B to activate the tick Upon activation a tick can be heard and a full bar graph i...

Страница 22: ...and visual alarms will activate when the preset limits are reached 15 When equipped with the percent volume sensor if the instrument encounters a gas it is not calibrated to it may read NSR or NSC fol...

Страница 23: ...e integrity of the O ring seals and connections on the probe and instrument If flow block can not be achieved contact the factory for assistance An air tight system is crucial for accurate readings Fr...

Страница 24: ...ay encounter NSR Non Standard Response or NSC Non Standard Combustible readings during the bar hole test Changing the GAS TYPE a selection in the USER MENU may help to identify a questionable vapor A...

Страница 25: ...nt will read in 1ppm increments up to 5000ppm auto range to LEL and then to v v To zero the instrument in the LS mode press and hold the SAVE ZERO C button until Autozero is displayed Any alarm can be...

Страница 26: ...ht side of the screen If O2 oxygen is not installed an X will appear Attach a purge probe Do not create a tight seal where the purge probe is inserted for sampling Allow for blow by so the unit does n...

Страница 27: ...he screen Press and release the TICK MENU B button The unit will exit back to the work display with the peak readings logged under the real time readings To reset the peak readings press and hold the...

Страница 28: ...on the instrument version A 180 second timed test will begin The time for this test is factory adjustable A countdown timer will show the remaining seconds of the test During this test period the det...

Страница 29: ...meet the specifications listed in this manual may require calibration or repair A calibration check does not update the calibration due date Full calibration is required to update these times Acalibr...

Страница 30: ...splayed using a 24 hour clock User adjustable Print Session Logs Cal Log access Smart Cal communication print CO test or print CF test is optional with some extended memory units Perform automatic tes...

Страница 31: ...use with calibration station Display future calibration due dates for each gas NOTE Standard LOG size is 6 events but up to 100 records can be stored per session BH Test CO Test and CF test NOTE These...

Страница 32: ...holding the TICK MENU B until the top line reads USER MENU Press release the SAVE ZERO BUTTON C until the bottom line displays SET CLOCK Press release the TICK MENU BUTTON B once to enter the menu Th...

Страница 33: ...ss release the SAVE ZERO BUTTON C until PRINT is displayed Press release the TICK MENU BUTTON B once to enter the menu Prepare the optional IR printer Aim the IR window on the right side of the instru...

Страница 34: ...nd a 45 60 second countdown on the bottom line The instrument will automatically check the LEL sensor and also the CO and H2S sensors if they are installed If each sensor tested reads at least 80 of t...

Страница 35: ...press release the TICK MENU BUTTON B to start the test A 20 second countdown will begin If the sensor shows proper depletion within this period PASSED will flash on the display Press release the POWER...

Страница 36: ...he gas type press release either button B or C You can select either NAT methane or PRO propane as your primary gas Once you have made your selection press release the POWER MUTE BUTTON A to store the...

Страница 37: ...is menu press release the POWER MUTE BUTTON A To Print a Calibration Log From the working display press hold the TICK MENU BUTTON B until the top line reads USER MENU Press release the SAVE ZERO BUTTO...

Страница 38: ...ill go back to the previous session The standard number of available stored sessions is factory set at 6 but is adjustable up to 100 To exit this menu press release the POWER MUTE BUTTON A to return t...

Страница 39: ...returns you back to the previous record Press the POWER MUTE BUTTON A to return to the USER MENU SMART CAL From the working display press hold the POWER MUTE BUTTON A for 2 3 seconds The display will...

Страница 40: ...the optional CF Test feature With this feature the instrument will automatically save the peak readings of all sensors while the unit is operating in the working display These peak readings are stored...

Страница 41: ...T will appear on the top line of the display Press release the SAVE ZERO BUTTON C until CF is displayed Press release the TICK MENU BUTTON B again and the instrument will auto zero and then enter the...

Страница 42: ...ed ppm CF reading and the calculated peak ppm CF level will be displayed on the right side of the screen If the proper conditions for an accurate test existed O2 below 18 9 the detected CO level calcu...

Страница 43: ...ER MUTE BUTTON A to return to the work display To Print a CF Test From the working display press hold the TICK MENU BUTTON B until the top line reads USER MENU Press release the SAVE ZERO BUTTON C unt...

Страница 44: ...OTE When calibrating the numbers shown on the display represent the numbers seen by the microprocessor and should not be confused with actual gas readings These readings will update every 5 seconds du...

Страница 45: ...ument to all gases detected by the sensors installed From the AUTO CAL menu press the TICK MENU BUTTON B Attach the prepared gas as listed on the display A countdown timer shows the time remaining to...

Страница 46: ...as Upon successful calibration the calibration due date reminder is automatically updated From this menu pressing the SAVE ZERO BUTTON C again will provide access to the SMART CAL automatic calibratio...

Страница 47: ...ntil the bottom line reads CAL Press release the TICK MENU BUTTON B once The bottom line will read AUTO CAL Scroll with the SAVE ZERO BUTTON C until CO 100PPM is displayed Apply 100ppm CO Air calibrat...

Страница 48: ...ON C until the bottom line reads H2S 25ppm Apply 25ppm H2S AIR to the instrument and press release the TICK MENU BUTTON B to start H2S calibration When the reading is satisfactory the display will fla...

Страница 49: ...TON C until the bottom line reads HCN 10ppm Apply 10ppm HCN N2 to the instrument and press release the TICK MENU BUTTON B to start HCN calibration When the reading is satisfactory the display will fla...

Страница 50: ...methane air calibration gas and press release the TICK MENU BUTTON B to start 50 LEL calibration When readings stabilize the display will read DATA SAVED indicating calibration is complete for that s...

Страница 51: ...L Press release SAVE ZERO BUTTON C to scroll until the bottom line reads 100 V V Apply 100 methane to the instrument Immediately press release the TICK MENU BUTTON B to start 100 methane calibration W...

Страница 52: ...LEL propane to the instrument Immediately press release the TICK MENU BUTTON B to start propane calibration When the reading is satisfactory the display will flash DATA SAVED indicating that calibrati...

Страница 53: ...A 20 second countdown will begin If the sensor shows proper depletion within this period PASSED will flash on the display Press release the POWER MUTE BUTTON A to return to the working display If the...

Страница 54: ...all date N COMP Specialized sensor compensation software CAL DUE REMINDER Alert system for calibration DUE ACK Requires operator to push a button when cal is overdue PROPANE 100 Requires calibration t...

Страница 55: ...Erase the AUTO LOG NG FACTOR Factor for methane content in 100 natural gas NSR Disable gas distinguishing software NSC Disable combustible inert identifier NSC LEL Concentration to activate NCS contro...

Страница 56: ...OLUTION 0 0 2 0 0 0 N COMP ON OFF OFF CAL DUE 30 45 60 90 180 360 DAYS 30 DUE ACK ON OFF OFF N2 FOR O2 ON OFF OFF SHOW SES ON OFF ON SHOW BH ON OFF ON SHOW AUTO ON OFF OFF ALARM LOW O2 17 5 20 5 19 5...

Страница 57: ...E TIME 0 120 SEC 10 SEC BH TIME 5 120 SEC 20 SEC ERASE AUTO ERASE AUTO PASSWORD REQ NG FACTOR 50 100 100 NSR ON OFF ON NSC ON OFF ON NSC LEL 1 0 10 0 2 0 AUTO BUMP 0 30 0 MUTE LATCH ON OFF OFF ERASE L...

Страница 58: ..._____________________________________________ ________________________________________________________ ________________________________________________________ ________________________________________...

Страница 59: ..._____________________________________________ ________________________________________________________ ________________________________________________________ ________________________________________...

Страница 60: ...shipment or failure which results from alteration tampering accident misuse abuse neglect or improper maintenance Proof of purchase may be required before warranty is rendered Units out of warranty wi...

Отзывы: