46
CALIBRATION
AUTO CAL (Continued)
Pressing the SAVE/ZERO BUTTON (C) will scroll through the individual
sensors that are configured in your G2 for calibration. Pressing the
TICK/MENU BUTTON (B) selects the option. Attach the proper gas and
wait for the instrument to provide a PASS or FAILED indication.
If the calibration has failed press the TICK/MENU BUTTON (B) to
attempt again or scroll to the next gas. Upon successful calibration the
calibration due date reminder is automatically updated.
From this menu pressing the SAVE/ZERO BUTTON (C) again will
provide access to the SMART-CAL automatic calibration system.
Pressing the TICK/MENU BUTTON (B) from SMART-CAL menu
begins the communication process. Consult the SMART-CAL system
instructions for connection details.
NOTE:
A calibration failure is indicated on the display by FAILED. Re-
calibration should be attempted. Any instrument that does not accept
calibration should be taken out of service.
Please contact SENSIT TECHNOLOGIES for any needed repairs.