Last updated 10/29/2020
16
Make sure that the 626 holder has a good vacuum. Finally, send message in slack in
#EMG channel: “Elsa is warming up now. V1 and V2 open, V3 and V4 of both Elsa and
626 are closed. 626 has a good vacuum and should be ready to go first thing in the
morning.”
b.
You are done after this step.
c.
If warming up the 626 holder, proceed to step 8.
8.
Once warm up is done, close V1.
9.
Open V3, then open V4 of holder #1.
10.
Open V3, then open V4 of holder #2.
11.
On Fridays, start the “ZEOLITE CYCLE” for 2 hours.
***If for any reason you cannot finish warming up the 626 and/or Elsa holder and open V3 and V4
to pump down the jacket of the holder by the end of your session, please send an email to
“626 and/or Elsa is still warming up right now. Only V1 and V2 are open, pumping down the tip
only. Can someone from EMG please close V1 and then open V3 and V4 first thing in the morning
to pump down the jacket of the holders?”
Содержание TF20