DreamStar Intro, Info, Auto patient manual
en GB – 33
Monthly
Device
•
Clean the outside of the unit using a damp tissue (cloth, paper towel) with a little water and one drop of
mild detergent.
•
Eliminate any trace of detergent in repeating this operation with a new tissue (cloth, paper towel)
moistened with water only.
•
Wipe the device thoroughly using a dry tissue (cloth, paper towel).
Air intake filters:
•
The high-efficiency filter, which is offered as an option, cannot be washed. It must be changed once a
month or more frequently if visibly dirty.
•
Change the filters systematically, as soon as they are torn or soiled or when the device displays a
reminder to do so.
Water chamber (if included):
•
After the water chamber has been disassembled and cleaned, you can let it soak for 15 minutes in a
solution of 1 part white vinegar to 9 parts water.
•
Rinse well with water to eliminate any trace of vinegar.
•
Allow to dry naturally.
WARNINGS
•
Do not use spray detergent. Harmful product residue could enter the air outlet, the air intake filter, or
the device interior, causing airway irritation.
•
Never use bleach at a concentration higher than 0.1%. For example: pour 200 ml of 2,6 % bleach into
5 liters of cold water.
•
Never use the device without making sure that an air intake filter is installed.
•
Check that the heating element is completely dry before reconnecting the device to the power supply.