Doc. No.: A05280692 Rev. AE ECO: 13200 Date: 11/04/2019
13
WARNING:
All life safety and electrical codes must be strictly followed.
As with any Infant Monitoring System, each application can be different. Use the guidelines below as a basic
understanding of what a standard application would be like.
1.
Seek prior approval from local life safety officials prior to installing an ID
locking system.
2.
Identify all equipment to be installed and inspect for any damage that may have resulted during shipment. If
damage is found notify the carrier immediately and arrange for inspection. Be sure to retain all shipping and
packaging materials.
3.
Install all communication wires from each exit system to the system annunciator location (computer location).
Secure Care strongly recommends that all wiring used to install Secure Care equipment be run at least 18 inches
from all other wiring and not be installed inside metal conduit. It is also strongly recommended to not utilize
cable trays used in networking installations and to secure the wires off of ceiling grids. If the wiring is installed
in any of these manners it could cause undesired induction of Radio Frequency (RF) interference with our
system.
4.
Determine the location of the exit control panel, making sure to accommodate for any local, stats, or federal
codes or guidelines including ADA (American Disability Act) requirements, and mount as required. Standard
applications would place this equipment on the wall at the center of the door height on the latch side.
5.
Mount the electromagnetic lock in strict accordance with the manufacturer installation instructions. Be sure to
comply with all Life Safety and Electrical Codes as required.
6.
Mount the magnetic door contacts provided with the Exit Kit on the latch side of the door. These contacts provide
a method of monitoring the open or closed status of the door.
7.
Route all necessary peripheral connection wires into the exit control panel mounting box. These would include
wires such as communication, magnetic contacts, exit power, magnetic lock power, etc.
8.
Prepare all wires and make connections to the exit system. Special care should be taken to prevent loose
connections and shorts.
9.
Proceed to the remote system annunciator location to mount as required. This device should be easily visible to
staff for monitoring of the system. Common locations are centralized nurse stations or staffed reception areas. .
10.
Make all necessary wiring connections as shown in
Section 9
of this manual.
11.
Plug in all power supplies and batteries.
Do not connect to a switched receptacle
.
12.
The system is now ready for tuning, programming, and testing.
13.
When using the STAT Cutband feature for an Adult application please visit and log into the Distributor portal
@
and refer to Technical Bulletin # 24
SECTION 4
TYPICAL INSTALLATION
Содержание KinderGUARD ID
Страница 8: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 8...
Страница 9: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 9...
Страница 10: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 10...
Страница 75: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 75...
Страница 77: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 77...
Страница 78: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 78...
Страница 81: ...Doc No A05280692 Rev AE ECO 13200 Date 11 04 2019 81...