System Operation Manual
SAT-DN-00434, Rev. B
LOBO
2017-05-16
Page 80 of 86
Copyright
2017 Sea-Bird Scientific. All rights reserved.
11.
Type
$BLS 1000
and press Enter to flush the bleach from the pump. Again, your typed characters
will not be echoed. This command will cycle the BLIS pump 1000 times to flush the system with
fresh water.
BLIS Pump: 1 of 1000
BLIS Pump: 2 of 1000
BLIS Pump: 3 of 1000
BLIS Pump: 4 of 1000
BLIS Pump: 5 of 1000
…
12.
When the purge completes, type
CONTROL-Z
to terminate the COMM program.
13.
Remove power from the STOR-X and disconnect the test cable. Prior to re-deployment,
remember to reset the “scheduled mode autorun” with the CONFIG program
!
14.
Remove the reservoir that is filled with fresh water; do not reinstall the bleach!!
15.
Reattach the flow cell and tubing.
11.5
WET Labs Cycle PO
4
Maintenance
Please refer to the Cycle PO
4
documentation.
11.6
Nortek Aquadopp Maintenance
Chapter 10 of the Aquadopp manual (RD5) contains maintenance tips. Key points:
•
Perform regular cleaning using a mild detergent, paying special attention to the transducers.
•
Remove any dirt from the pressure sensor holes.
•
Lubricate the male connector pins with the lubricant provided in the Aquadopp shipping case.
•
Antifouling paints can be used.
•
Replace internal desiccant annually.
•
When used with LOBO, the Aquadopp uses external power, so there is no reason to replace the internal
batteries. You may wish to remove the internal battery pack prior to deployment, please refer to the
user manual for instructions.
11.7
Nortek Continental Maintenance
Chapter 6 of the Continental Manual (RD6) contains maintenance tips. Key points:
•
Perform regular cleaning using a mild detergent, paying special attention to the transducers.
•
Remove any dirt from the pressure sensor holes.
•
Lubricate the male connector pins with the lubricant provided in the Continental shipping case.
•
Antifouling paints can be used.
•
Replace internal desiccant annually.
11.8
WET Labs ECO-FLNTUS Maintenance
The ECO-FLNTUS manual includes maintenance information. Key points:
•
Warning!
Do not use acetone or other solvents to clean the sensor!
•
Do not move the Bio-Wiper with your finger
–
this can damage the wiper motor and void the warranty
•
Rinse with fresh water after each use. Pay particular attention to the sensor face. Use soapy water for
grease or oil, and a soft cloth.
•
Avoid scratching the sensor face.