System Operation Manual
SAT-DN-00434, Rev. B
LOBO
2017-05-16
Page 58 of 86
Copyright
2017 Sea-Bird Scientific. All rights reserved.
ISUSSchedule is presently only available as a MS-Windows application and so can not reside on the Linux
server. It requires MS-Windows NT, 2000, or XP to run. The program is provided as a self-installing
executable. In order to execute the program, the MatLab Component Runtime has to be installed. This
component is present with all installations of MatLab 7 (R14). For users without a MatLab license, a stand-
alone runtime must be installed.
A simple manual is provided with the ISUSSchedule program, however a thorough procedure is provided
below to generate a common schedule for a typical LOBO system. In this example, we have the following
physical configuration:
Port 1: WQM
Port 2: ISUS
Port 3: ECO-CDS
Port 4: Unused
Port 4: Unused
We wish to optimize the schedule to maximize life of the battery pack while maintaining good temporal
resolution. A once an hour schedule is acceptable. The WQM must be turned on early to ensure that the
sampling volume is well flushed and to allow time for the dissolved oxygen sensor to stabilize; about 1
minute should suffice. Because the ISUS is plumbed with the WQM sensor, it should acquire data during
the same time. One minute of acquisition time should be sufficient. To eliminate any chance of interference
between the ECO-FLNTUS in the WQM and the separate ECO-CDS, the ECO-CDS power-up and acquisition
will occur separately; 20 seconds of data for this sensor should suffice.
Note that this example may seem quite a bit more complicated than a typical STOR-X schedule. However,
these sensors have been extensively tested at Sea-Bird Scientific, and this schedule is highly optimized for
performance.
1.
Start the ISUSSchedule program. You will see a DOS window open, followed by the actual
Windows application.
2.
In the Instrument drop down list, select STOR-X as shown below.
3.
As explained above, only Ports 1-3 and the modem port
(don’t forget this!!)
are required, so select
these as shown below.