Section 4: Setting Up MicroTSG
SBE 45
30
Polled Sampling
Commands
These commands are used by an external controller to request a sample from
the MicroTSG. The MicroTSG stores data for the most recent sample in its
RAM. The MicroTSG does not automatically go to sleep after executing
these commands. Do not send these commands if the MicroTSG is sampling
data at pre-programmed intervals (defined by
Interval=
and
SingleSample=
).
TS
Take sample, hold converted data in RAM,
output converted data.
TSR
Take sample, hold raw data (temperature and
conductivity only) in RAM, output raw data.
SLT
Send last sample from RAM, output converted
data, then take new sample and hold converted
data in RAM.
TH
Take sample, hold converted data in RAM.
SH
Send held converted data from RAM.
Testing
Commands
TT
Measure temperature 100 times or until Esc
key is pressed, output converted data.
TC
Measure conductivity 100 times or until Esc
key is pressed, output converted data.
TTR
Measure temperature 100 times or until Esc
key is pressed, output raw data.
TCR
Measure conductivity 100 times or until Esc
key is pressed, output raw data.
35
Содержание SBE 45 MicroTSG
Страница 2: ...This page intentionally left blank ...
Страница 4: ...Manual Generation Date 13 February 2012 1 ...
Страница 6: ...This page intentionally left blank 3 ...
Страница 7: ...4 ...
Страница 98: ...CALIBRATION SHEETS SBE 45 Temperature Calibration S N 0402 1 SBE 45 Conductivity Calibration S N 0402 2 95 ...
Страница 112: ...109 ...
Страница 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Страница 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Страница 129: ...126 ...
Страница 130: ...127 ...
Страница 131: ...128 ...
Страница 132: ...129 ...
Страница 133: ...130 ...
Страница 134: ...131 ...
Страница 135: ...132 ...
Страница 136: ...133 ...
Страница 137: ...134 ...
Страница 138: ...135 ...