Section 4: Setting Up MicroTSG
SBE 45
23
4.
In the Communications menu, select
Options / Cycle baud
when connecting
.
5.
Click Connect on the Toolbar. SEATERM tries to connect to the
MicroTSG at the baud set in Step 3. If it cannot, it cycles through all other
possible baud rates to try to connect. When it connects, the display looks
like this:
S>SBE45 V 1.1b
S>
This shows that correct communications between the computer and
MicroTSG has been established. If the system does not respond as shown:
•
Click Connect again.
•
Verify the correct instrument was selected in the Configure menu and
the settings were entered correctly in the Configuration Options
dialog box.
•
Check cabling between the computer and MicroTSG.
6.
Display MicroTSG status information by clicking Status on the Toolbar.
The display looks like this:
SBE45 V 1.1b SERIAL NO. 1258
not logging data
sample interval = 30 seconds
output conductivity with each sample
do not output salinity with each sample
do not output sound velocity with each sample
do not start sampling when power on
do not power off after taking a single sample
do not power off after two minutes of inactivity
A/D cycles to average = 4
7.
Command the MicroTSG to take a sample by typing
TS
and pressing the
Enter key. The display looks like this (if
output conductivity with each
sample
,
do not output salinity with each sample
, and
do not output sound
velocity with each sample
displayed in response to the status command in
Step 6) :
23.7658, 0.00019
where
23.7658 = temperature in degrees Celsius
0.00019 = conductivity in S/m
These numbers should be reasonable; i.e., room temperature and
zero conductivity.
8.
Establish the sampling scheme, after reviewing the information in this
section on sampling modes and commands.
9.
Command the MicroTSG to go to sleep (quiescent state) by typing
QS
and pressing the Enter key. The response indicates whether the PCB’s J1
jumper is in the Normal or Autopower configuration:
•
Autopower – system returns
S>
prompt.
•
Normal – system does not return
S>
prompt.
If necessary, remove the PCB and move the jumper to the desired pins.
The MicroTSG is ready for deployment.
Notes:
•
See
Power-Up Jumper Check
in
Section 3: Installing System
for a description of Autopower
and Normal.
•
See
Appendix II: Electronics
Disassembly/Reassembly
to access
the PCB and move the jumper.
28
Содержание SBE 45 MicroTSG
Страница 2: ...This page intentionally left blank ...
Страница 4: ...Manual Generation Date 13 February 2012 1 ...
Страница 6: ...This page intentionally left blank 3 ...
Страница 7: ...4 ...
Страница 98: ...CALIBRATION SHEETS SBE 45 Temperature Calibration S N 0402 1 SBE 45 Conductivity Calibration S N 0402 2 95 ...
Страница 112: ...109 ...
Страница 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Страница 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Страница 129: ...126 ...
Страница 130: ...127 ...
Страница 131: ...128 ...
Страница 132: ...129 ...
Страница 133: ...130 ...
Страница 134: ...131 ...
Страница 135: ...132 ...
Страница 136: ...133 ...
Страница 137: ...134 ...
Страница 138: ...135 ...