39
Indication
Error description
Error: Code mismatching Error
The inserted test strip and codechip do not match.
Solution
Change the appropriate test strip or codechip.
Indication
Error description
Error: Flap Open
The chamber flap of the Analyzer is open during measurement.
Solution
Close the chamber flap and do not touch or open the flap during measurement.
Indication
Error description
Error: Blood Error
Measurement chamber flap is closed for testing without blood sample.
Solution
Re-test with sufficient blood sample.
Indication
Error description
Error: Communication Error with codechip
The Analyzer cannot communicate with the inserted codechip.
Solution
Re-insert the codechip. If the error message persists, please contact your local distributor.
Indication
Error description
Error: Internal Error
There is an internal error of the analyzer.
Solution
Turn the Analyzer off and then back on again. If the error message persists, please contact your local distributor.