38
Indication
Error description
In case of lipid test strip
In case of glucose test strip
Error: blood sample Error
An insufficient amount of blood was applied.
Solution
Discard this test strip and test again using a new test strip with a larger sample.
Indication
Error description
Error: Expired strip (Only for a Lipid test strip)
The Lipid test strips are expired.
Solution
Discard this test strip and test again using a new lipid test strip that has not expired. If the strips are not expired
and you still get E-3 then this means the year and/or date are not set correctly. Please refer to the section on
setting the Analyzer.
Indication
Error description
Error: Temperature Error
If the environmental temperature is above or below the operation range of the Analyzer, a thermo Analyzer icon
will appear on the display.
Solution
Move to an area between 18-32°C(64-90°F) for lipid or 10-45°C (50-113°F) for glucose, wait for 30 minutes, and
perform a test. Do not artificially heat or cool the Analyzer.
Indication
Error description
Error: Communication Error
The communication between Analyzer and computer or printer is failed.
Solution
Connect the Analyzer and external device again.
If the error message persists, please contact your local distributor.