background image

23

CHAPTER 3. 
Using the Analyzer Memory

1. Memory Map

The MultiCare Analyzer can store 999 test results with date, time, and flags 
to  let  you  review  them  in  order.  If  the  memory  is  full  and  a  new  result  is 

added, the analyzer deletes the oldest result. 

Data

{ index=2}

Oldest Data

{ index=999}

Left Button

Latest Data
{ index=1}

Strip wait

Data

{ index=998}

Latest Data

{ index=1}

Right Button

Oldest Data

{ index=999}

Strip wait

 

Содержание MultiCare 03MA10

Страница 1: ...operate Please read this User Instruction Guide carefully before using the analyzer Familiarize yourself with the required preparations and the measurement procedure before performing the first measur...

Страница 2: ...iagnostic medical device This system is intended to use outside the body in vitro diagnostic use Manufacturer Reference number Date of manufacture To indicate the date of manufacture for this analyzer...

Страница 3: ...alues 24 3 Delete Data Stored in Memory 24 CHAPTER 4 PRINTING AND DATA TRANSFER 25 1 Printing 25 2 Data Transfer 26 3 Data transfer via Bluetooth 26 CHAPTER 5 CONTROL TEST 28 CHAPTER 6 CHECK STRIP TES...

Страница 4: ...the optic sensor to measure the reflected light from the membrane of the test panel as the quantity source The specimen from human should be processed for the preparation using the components of the M...

Страница 5: ...re System is only for in vitro diagnositc use Place the analyzer on a level surface Strong electromagnetic fields may impair the function of the analyzer Do not use the analyzer close to sources of st...

Страница 6: ...8 Overview of MultiCare Analyzer A B C D E F H I G J K L M...

Страница 7: ...anel Slot Part for inserting a test panel into an analyzer G LCD Display Shows test results messages and test results stored in memory H Battery compartment lid Provides access to the battery compartm...

Страница 8: ...cate pressing the start button Indicate the operating temperature is out of the range Unit of a test result for all parameter Indicate the time format Capacity of batteries Indicate the connection wit...

Страница 9: ...n the memory With a set of fresh batteries you will normally be able to perform at least 200 measurements When the battery warning is displayed for the first time approximately 10 measurements can sti...

Страница 10: ...you can hold down the Enter button the next time you switch on the analyzer The display screen is shown during the time while the button is pressed After inserting or replacing batteries confirm that...

Страница 11: ...r turning an analyzer on or at the standby state If you want to leave the setting mode this is only possible if a date and time setting is made press the SET PRT button shortly 2 If the displayed sett...

Страница 12: ...the beep off feature no sound will be made Stage 2 Setting Year In the second stage you set the Year in the Date Time setting After setting Beep the display for setting the Year will appear Set the c...

Страница 13: ...te Time setting The analyzer can display the time in either 12h format or 24h format Set the preferred format on the display by pressing either the left or right button and select by pressing the Ente...

Страница 14: ...ff by pressing either the left or right button and then select the preferred feature by pressing the Enter button NOTE The Auto Printing function is for printing the result right after testing Stage 8...

Страница 15: ...ther the left or right button and select by pressing the Enter button Stage 10 Setting eAG Unit In the tenth stage you set the unit of the estimated average of the blood glucose Set the preferred unit...

Страница 16: ...lly via Bluetooth After the setting of lipid unit the Bluetooth setting will appear on the screen Select it on or off by pressing the left or right button and then save it by pressing the Enter button...

Страница 17: ...P 3 Insert a Code Chip Coding The code chip is included in the test kit package The code chip provides the analyzer with important information on the production specific properties of each test kit in...

Страница 18: ...tion ALWAYS Read the appropriate test kit insert Place the analyzer on a level surface Keep the surrounding test site area tidy and clean Make sure that all display elements are displayed NEVER Never...

Страница 19: ...ssay 1 Insert the test panel into the MultiCare analyzer When the test panel reaches the correct position panel is automatically inserted deeper to the analyzer If the test panel is inserted appropria...

Страница 20: ...y 4 Press the start button The result will be displayed on the screen in 3 or 5 minutes 5 When the test is done pull out the used test panel The analyzer shuts off automatically after you remove the t...

Страница 21: ...with date time and flags to let you review them in order If the memory is full and a new result is added the analyzer deletes the oldest result Data index 2 Oldest Data index 999 Left Button Latest Da...

Страница 22: ...mber of stored results is 999 3 Delete Data Stored in Memory You can delete data while reviewing stored memory items To delete the stored data press the Enter button for three seconds when accessing t...

Страница 23: ...nction on in the setting mode the analyzer will not display the Printing symbol When a test is done the test result sheet is automatically printed without any action Manual Printing If you press the S...

Страница 24: ...MA20 model transfers the data from an analyzer via Bluetooth The model number is indicated on the label of the analyzer Data Transfer The data measured by an analyzer can be transferred to PC via Blue...

Страница 25: ...appear on the screen At the same time the analyzer will display OK message on the screen If you receive the both messages on the PC and the analyzer pairing is done successfully NOTE The xx xx xx xx x...

Страница 26: ...ing properly you should regularly use the control materials for a control test Make a habit of carrying out the control test during the following situations When you open a new test kit After replacin...

Страница 27: ...new control tablet or warm up cool down to room temperature Did you insert the test kit firmly into the analyzer Make sure the test panel is inserted into the test panel slot until it will go no furt...

Страница 28: ...CAUTION The check strip test does not replace control test How to use the Check Strip There are two kinds of check strips white and gray The MultiCare Analyzer can read the light reflectance value fo...

Страница 29: ...ct the strip Scan CHE strip If measured value iswithin the range OK If measured value is not within the range EEE White calibration Panel Wait Enter Button Press Enter and Start buttons at the same ti...

Страница 30: ...egularly and immediately after it becomes dirty Always switch off the analyzer before cleaning it Use only the following items for cleaning Ordinary lint free cotton buds Ordinary lint free tissues Or...

Страница 31: ...it is not giving accurate results make sure that your test kits and control sets haven t expired and then run the control test or check strip test Analyzer Keep the test panel slots free of dust Prot...

Страница 32: ...diately 2 Error Messages In certain circumstances error messages may appear on your display Generally you should first try the solutions suggested for the respective error If the problem persists plea...

Страница 33: ...n will appear on the screen Solution Move to an area with temperatures between the applicable temperature range and perform a test Do not artificially heat or cool the analyzer Communication Error The...

Страница 34: ...contact your local customer support and service center Communication Error with the Code Chip The code chip is broken So the analyzer could not communicate with the code chip Solution Re insert the c...

Страница 35: ...zer should be used with the appropriate MultiCare Test Kit 8 Do not use the MultiCare Test Kit if its code number does not match with the number of the code chip that is inserted into the analyzer 9 A...

Страница 36: ...alyzer meets the electromagnetic immunity requirements The chosen basis for electrostatic discharge immunity testing was basic standard IEC 61000 4 2 In addition it meets the electro magnetic emission...

Страница 37: ...1556beon gil Yeongtong gu Suwon si Gyeonggi do 16690 REPUBLIC OF KOREA Manufacturing Site 74 Osongsaengmyeong 4 ro Osong eup Heungdeok gu Cheongju si Chungcheongbuk do 28161 REPUBLIC OF KOREA www sdb...

Страница 38: ...MEMO...

Отзывы: