28
CHAPTER 5.
Control Test
It is important to perform Control Tests with more than one level of control
materials to assure your system is working properly.
Why You Should Do The Control Test:
•
The control material is used to check that the analyzer and the test
kits are working together as a system and you are performing the test
correctly.
•
It is very important that you perform this simple check routinely to make
sure you get an accurate result.
When do Control Test:
To ensure that the analyzer is functioning properly, you should regularly use
the control materials for a control test. Make a habit of carrying out the
control test during the following situations:
•
When you open a new test kit
•
After replacing the batteries
•
After cleaning the analyzer
•
When you doubt that the measured values are correct
The control test is performed in the same way as a regular measurement
with the difference that control are used instead of sample.
Materials required for Control Test
•
MultiCare Analyzer
•
MultiCare Test Kit
•
Control material for the parameter you want to check
Performing a Control Test
NOTE
Please refer to the control insert of each compatible control
materials separately provided for detailed test procedure.
Содержание MultiCare 03MA10
Страница 6: ...8 Overview of MultiCare Analyzer A B C D E F H I G J K L M...
Страница 38: ...MEMO...