Troubleshooting for Control Test
Check
Action
Did you perform the test while in
the control mode?
If not, do the test again. Insert a test panel .
Press the left (
) button for three seconds
to display the control icon.
Did you see “control icon” on the
screen with the result?
Have the test kits and/or control
expired?
Make sure that the test kits and control are
not past their expiration dates.
Are the testing conditions suitable
for MultiCare test kit?
If not, retest with the new control tablet, or
warm up/cool down to room temperature.
Did you insert the test kit firmly into
the analyzer?
Make sure the test panel is inserted into the
test panel slot until it will go no further.
Did you follow the procedure
correctly?
Read the instructions for use again and
retest.
Were the test kits stored correctly?
If not, retest with a new test kit.
Is the analyzer damaged? Does it
show an error code?
If yes, contact your local distributor.
Is the control result outside the
assignment range?
Repeat the test. If you get the same results,
do not use your analyzer and test kit until
you solve the problem.
If you still have problems, contact your local
distributor.