Version 3
28
|
Mini
WAVE
Modual Indicator
The information of this document pertains to firmware version 3.0.0 and up
Symbol conventions
A 7-segment display has the following descriptors:
To describe the animation pattern, the
segment sequence will be written as
“A-B-C-D-E-F” for example: this pattern
is repeated during the ramp stage.
System booting
When the power is applied to the system, the letters H,E,L,L,O are displayed as a segment test indication. Those letters use all indicator
segments so if this message is not seen at power-up, there could be a segment malfunction. There is no self-detection for correct segments
drive so the user only can see an indicator fault.
After this message, the system tests many hardware peripherals and the following error messages could be seen:
•
LE: Line frequency Error -> there is no AC signal activity detected
•
CE: Calibration Error -> Calibration tables are empty or inconsistent
•
tE<n>: Thermopile Error for location <n> -> Data read is inconsistent
•
HE1: Home Error 1 -> Exhaust Door Homing failed
•
HE2: Home Error 2 -> Rack Rotation Homing failed
This is followed by the firmware version (3 digits)
Finally the Unit Number is displayed and stays until a digestion starts.