Version 3
22
|
Diagnostic Tab
CommView is used to validate the
communication between controller and
module
Operations is only available to service
personnel. This tab is not functional if a
digestion is in progress
Refresh Legend is used to pull all of
the communication protocols from the
modules to help debug a problem. This
will be used by factory trained service
people to diagnose problems on the
Module
View and Stop is to suspend the
acquisition of data
Operation
The commands on this tab are to be
used by factory trained personnel only
with the exception of reading the IR
temperature sensor when the unit is not
performs a digestion
XB Speed Control is used to run the
exhaust fan. Programmed in percent of
full speed
PID Control is used to query the current
PID values
Control/Read Filament is used to query
the magnetron filament power
Control/Read Fans is used to turn the
fans on for testing after servicing
Calibration/PID is used for debugging in
manufacturing
Cool Jet is used to turn the cool jet On
and Off for testing