Troubleshooting
Safety Messages
Safety messages may occur during system operation due to a variety of conditions as included below. For further assistance, consult your lab
manager or
SCP SCIENCE
Service Department or your local distributor.
LIST OF WARNING AND SAFETY MESSAGES
Message
Description
Action
1
Module-1 USB was unplugged.
Either the USB cable was physically
unplugged or Module-1 has a
problem.
Stop digestion. Try to re-run.
2
Module-1 has Calibration Table Error.
Calibration table is either corrupted
or blank.
Try to send calibration table from screen
to module.
3
The line voltage of Module-1 is out of range.
Power line has a problem.
Digestion stops. Check power line voltage.
4
The line frequency of Module-1 is out of range.
Power line has a problem.
Digestion stops. Check power line
frequency.
5
Module-1 has Filament problem.
Digestion stops.
6
Module-1 has High Voltage problem.
Digestion stops.
7
Module-1 has IR Sensors COM problem.
Module-1 has internal
communication problem.
Digestion stops.
8
Module-1 has IR Sensors Error.
Problem of Infrared sensor.
Digestion stops.
9
Module-1 detected excessive vibration.
Digestion stops.
10
Module-1 magnetron overheated.
Digestion stops.
11
Module-1 digestion timed out.
Digestion stops.
12
Module-1 temperature is too high for tube-2.
Digestion stops.
13
Module-1 temperature is too low.
Digestion stops.
14
IR communication problem for Channel #2.
Channel #2 has IR communication
problem.
User decides to continue or stop digestion.
15
IR Sensor error for Channel #2.
Channel #2 has IR sensor error.
User decides to continue or stop digestion.