Version 3
|
17
Methods Tab
This tab lists all of the factory and user programmed methods. Highlight a programmed method and then select one of the active action
buttons at the bottom.
Name is the method selected from the
left hand side list. Some methods are
factory programmed and cannot be
deleted but may be copied and modified
as required.
Created Time is a time stamp from when
the method was saved
Last Modification is the time stamp
after any modifications were made and
re-saved
Description is extra information that was
added during the method creation on
the methods page.
Press New to create a new digestion
procedure
Delete is used to delete the method
completely. This can only be done in the
Service and Administrator access levels.
Copy allows a factory programmed
method or a user method to be copied.
The method name must be changed to
be saved
Edit allows a user defined method to be
modified when done in the service or
administrator access level
View allows any operator to view the
parameters in any method
Save to USB save the method to an installed
USB stick on the back of the controller.
Name is a required field to be filled out
as method are stored by name.
Description is a field for extra
information. This information is stored
and timed stamped until the method is
completely deleted. Modifications can be
made and the time stamp will updated.
CAUTION
WARNING
The maximum recommend temperature for digestions is
200 °C. This is for all vessels Teflon
®
and quartz vessels. If
this temperature is exceeded there is a highly possibility of
venting due to pressure achieved in by using mineral acids
in a digestion. HCl and H
2
O
2
create more pressure at lower
temperatures.
Warranty is not covered due to venting when method
temperature exceeds 200 °C.