Version 3
12
|
Module Tab
Select the Method from the pull down
tab under method. The system will use
the parameters defined in the method to
digest the samples. The system displays
the Module (Module -1) which is being
controlled. The module itself will have
a flashing number on the front of the
system to indicate which module is being
operated. Vessel Type must be selected
from the list, Quartz for quartz digestion
vessels and Teflon for Teflon
®
vessels in
quartz sleeves. This ensures the correct
temperature algorithm is used to control
the digestion temperature.
During the digestion, each of the vessels are being monitored for temperature and the average sample temperature is used to control the
Mini
WAVE
. If there is a significant temperature difference for one vessel either higher or lower than the other vessels for a certain time, the
system will stop the digestion and identify the suspected position.
Re-Scan is used to update the controller
when new modules are connected. The
controller can monitor and control up to 4
modules simultaneously. This only needs
to be done at initial installation and after
any disconnection of the module. The
controller assigns module numbers which
are then accessed in the Reports and
Utilities tabs.
Once the method and vessel type has
been selected, place the rack in Cavity,
close the lid and press Start Digestion.
To maneuver from Module to Module
press the Module-1 button once to
move to Module-2 and twice to move to
Module-3. It will only maneuver between
the modules that are connected.
Rack Name maybe modified by
highlighting the cell and type the new
name. The keyboard will appear once the
cell is accessed.