OPERATIONAL PRECAUTIONS
Digi
PREP
block must be connected to a
controller that is properly grounded. In the event
of an electrical short circuit, grounding reduces
the risk of electrical shock by providing an escape
wire for electrical current.
Digi
PREP
controllers are
supplied with an electrical cord that has a grounding
wire with a grounding plug. Ensure that the plug
is plugged into an outlet that has been properly
installed and grounded. Consult a qualifi
ed
electrician if there is any doubt about the existence
of proper grounding at the outlet.
Digi
PREP
block is designed to be lifted from
the bottom only. Ensure that the controller is
disconnected prior to moving the block. Do not
attempt to lift the block by the top white trim
as damage may result.
Digi
PREP
blocks are elevated-temperature
digestion systems with exposed hot surfaces. Be
careful around the instrument when it is in operation.
SAFETY INFORMATION
A minimum of safety considerations must be
followed when working with
Digi
PREP
low
temperature blocks in order to maintain good
laboratory practice:
• Do not place the
Digi
PREP
blocks on a surface
containing fl ammable material.
•
Digi
PREP
block requires a clearance of three
inches (7.5cm) on all sides.
•
Digi
PREP
block should be operated in a fume
hood.
• Do not place the
Digi
PREP
controller in a fume
hood.
•
Digi
PREP
system must be plugged into a
three-prong grounded outlet
Содержание DigiPREP
Страница 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Страница 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Страница 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Страница 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Страница 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...
Страница 86: ...SCP SCIENCE 86 KeyPad Version 1 0 NOTES ...
Страница 87: ...KeyPad Version 1 0 SCP SCIENCE 87 NOTES ...