SCP SCIENCE
28
KeyPad
Version 1.0
THE
Digi
PREP
BLOCK IS NOT HEATING UP
DigiPREP connection verifi cation
• Ensure that the cable between the KeyPad
controller and the heating block is correctly
inserted on both sides.
• Verify that the voltage selector of the heating
block is properly set.
COM2 dongle verifi cation
When the
COM2
of the KeyPad
controller is not connected to a
Digi
SET
system, the port must be
shorted.
• Make sure that sections 1 and 2 have been verifi ed
before continuing.
• Ensure that the
Digi
SET
dongle is fully inserted
into
COM2
port (see fi gure 2)
• If so, before opening the bottom plate, verify with
a multimeter that the
COM2
dongle is shorted. If
not, replace the part with a new one.
• If the dongle is ok, remove the bottom by
removing the 4 screws that hold the feet in place.
Check if the
Digi
SET
cable (see fi gure 15 for
position) is fully inserted.
FIGURE 15:
DIGI
SET
CONNECTOR LOCATION
Legend
Digi
SET
connector
INTERNAL WIRING CONNECTION VERIFICATION
• After removing the bottom plate, verify the
connectivity between the heating block external
connector and the PCB (see fi gure 16).
FIGURE 16: HEATING BLOCK CONNECTOR LOCATION
Legend
1. Heater connector (Hot heating block)
2. Neutral connector (Neutral heating block)
3. Earth Ground
DIGISET VERIFICATION (ONLY IF APPLICABLE)
Make sure that the
Digi
SET
is fully fonctional (see
NOTE
Содержание DigiPREP
Страница 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Страница 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Страница 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Страница 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Страница 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...
Страница 86: ...SCP SCIENCE 86 KeyPad Version 1 0 NOTES ...
Страница 87: ...KeyPad Version 1 0 SCP SCIENCE 87 NOTES ...