Next, ensure that
Digi
PREP
controller is set
to the correct settings for controlling
Digi
PREP
block in use. (
HT
for all of
Digi
PREP HT
high
temperature digestion blocks) With the controller’s
power switch in the “off” position, verify the following:
• Voltage selector switch on the back of the
controller is properly set to 230 volts.(The Colour
Touch Screen & KeyPad do not have a voltage
selector
)
• Connect the blocks umbilical cord to the
controller.
Disconnecting or reconnecting the
umbilical cable while the controller is
turned on, especially while it is
on
the
RUN
screen, will likely result
in permanent damage to the
instrument.
• Connect the power cord to the back of the
controller.
• Connect the
Digi
SET
dongle to the back of
thecontroller.
Digi
SET
is not typically used with
the
Digi
PREP HT
Digestion blocks.
The system is now ready to be
plugged in and turned on. Consult
the controller manual for
programming and operation
instructions.
DUAL PROBE OPTION
The
dual
PROBE
may be retrofi tted in the fi eld
or may have come factory installed. This option
allows the lab to verify the
Digi
PREP HT
block
temperature independently of the
Digi
PREP
Controller used. The dual
PROBE
contains two
separate RTDs in one sheath; one sensor is
connected through the umbilical cord to the
Digi
PREP
controller to monitor and control the block
temperature. The second RTD
PROBE
is connected
to the 2 pin connector on the back of the
Digi
PREP
HT
system to allow for the monitoring of
the precision of block controller display temperature.
(Or the correct size hole may be made to house the
panel mount connector for fi eld installation). The 2 pin
female connector on the back of the
Digi
PREP
HT
system is supplied with a mating cable with
two bare ends. This cable is to facilitate the
connection of the second
PROBE
sensor to an
instrument capable of reading a 100
RTD. If
an Ohm meter is being used, the typical value at
100°C is 138.5
, and for 380°C the resistance
value is 240.13
(see table page 61.) It is
recommended to use a certifi ed controller capable
of measuring a temperature from 100 ohms platinum
RTD. (0.00385 factor)
The tolerance of the temperature
PROBE
s used
on the
Digi
PREP HT
systems is of a ‘Class A’
specifi cations. A typical variation at 100°C is ±
0.15°C and at 400°C is ± 0.95
o
C.
Parts for the Dual
PROBE
Option (fi eld installation)
complete with mating cable.
NOTE
WARNING
Field installation kit
Mating Cable
Содержание DigiPREP
Страница 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Страница 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Страница 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Страница 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Страница 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...
Страница 86: ...SCP SCIENCE 86 KeyPad Version 1 0 NOTES ...
Страница 87: ...KeyPad Version 1 0 SCP SCIENCE 87 NOTES ...