SCP SCIENCE
14
KeyPad
Version 1.0
OPERATION PARAMETERS
Return to the menu screen. Select OPERATION
and press enter.
The next screen will prompt you to enter the target
temperature:
Using the number keys, enter the desired operating
temperature. Press enter.
The next screen will prompt you to enter the target
time:
The target time is the duration for which you want
the system to hold the target temperature. If the
alarm is activated, it will also use this time. The alarm
will sound at the end of the time sequence. Enter
the desired target time (in minutes) using the number
keys and press enter to accept the selection.
From the next screen, you can choose whether or
not you want the heater to remain on at the end of
the target time.
Select OFF to turn the heater off at the end of the
target time, and ON if you want the block to remain
at the target temperature indefi nitely. (Press enter to
accept the selection)
RUN SCREEN
Once you are fi
nished entering the operation
parameters, the run screen will be displayed:
The run screen displays the current temperature
and time remaining. The temperature is displayed
to the nearest 0.1 of a degree, and the time is
displayed in minutes and seconds.
An asterisk will appear occasionally in the upper
right corner of the display. This indicates that the
heating element is fi ring.
If "AL" is displayed in the bottom right corner of the
screen, the alarm is set to sound when the remaining
time runs out. Pressing the alarm key will toggle the
alarm off. You can also use the alarm key to silence
the alarm.
To reset the timer, press the target time key to
return to the target time screen.
CALIBRATION SCREEN
Selecting CALIBRATE from the utilities screen will
bring you to the calibration screen, which is used to
enter calibration infromation for the
Digi
PROBE
.
Calibration is done by comparing the keypad
display temperature to an NIST-calibrated (or similar)
thermometer. The difference between the reading
from the calibrated device and the keypad display
temperature will be entered into the system.
Not all thermometers have the same
response
time.
Run the system using a
Digi
PROBE
inserted
in a
Digi
TUBE fi lled with water. On the keypad,
press the target temperature key and set the target
CALIBRATION
#POINTS
0 (0-5)
MENU
OPERATION UTILITIES
TARGET TEMPERATURE?
25 TO 180
o
C
XXX.X
TARGET TIME?
0 TO 999
XXX
END OF TIME?
HEATER OFF
ON
TEMPERATURE 95.0
o
C
TIME
120:23
AL
NOTE
Содержание DigiPREP
Страница 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Страница 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Страница 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Страница 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Страница 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...
Страница 86: ...SCP SCIENCE 86 KeyPad Version 1 0 NOTES ...
Страница 87: ...KeyPad Version 1 0 SCP SCIENCE 87 NOTES ...