OPERATIONAL PRECAUTIONS
Digi
PREP HT
must be grounded. In the event
of an electrical short circuit, grounding reduces the
risk of electrical shock by providing an escape wire
for electrical current.
Digi
PREP HT
is supplied with
an umbilical cord that has a ground wire that is
connected through the controller to earth ground.
Ensure that the plug is plugged into an outlet that
has been properly installed and grounded. Consult
a qualifi ed electrician if there is any doubt about the
existence of proper grounding at the outlet.
Digi
PREP HT
is designed to be lifted from
the bottom only. Ensure that the controller is
disconnected prior to moving the system. Do not at-
tempt to lift the system by the top white trim or the ‘U’
channel as damage may result.
Digi
PREP HT
is an elevated-temperature
digestion system with exposed hot surfaces. Be
careful around the instrument when it is in operation.
Digi
PREP HT
digestion block should be
operated in a fully functional fume hood.
SAFETY INFORMATION
A minimum of safety considerations must be
followed when working with the
DigiPREP
HT
in
order to maintain good laboratory practice:
• Do not place the
DigiPREP
HT
on a surface
containing fl ammable material.
•
Digi
PREP HT
requires a clearance of 3 inches
on all sides.
•
Digi
PREP HT
must be operated in a fume hood.
• Do not place the
Digi
PREP
controller in a fume
hood.
•
Digi
PREP HT
is connected to a
Digi
PREP
controller.
•
Digi
PREP HT
requires 230V ± 10% three-prong
grounded
outlet.
Содержание DigiPREP
Страница 61: ...KeyPad Version 1 0 SCP SCIENCE 61 ELECTRICAL SCHEMATIC DigiPREP HT ...
Страница 73: ...KeyPad Version 1 0 SCP SCIENCE 73 HEATER VERIFICATION ...
Страница 83: ...KeyPad Version 1 0 SCP SCIENCE 83 ...
Страница 84: ...SCP SCIENCE 84 KeyPad Version 1 0 ...
Страница 85: ...KeyPad Version 1 0 SCP SCIENCE 85 ...
Страница 86: ...SCP SCIENCE 86 KeyPad Version 1 0 NOTES ...
Страница 87: ...KeyPad Version 1 0 SCP SCIENCE 87 NOTES ...