○ Maximum Interval: 24 hours.
MEMORY MODE
•
Circular Me
mory:
Allows you to select so that the devic
e continuously logs in memory. When the memory is full, the oldest data will be erased
so that the most r
ecent data can be saved.
•
Me
mory Full:
All
ows you to select the device to pause records when the internal memory has no more available space.
START MODE
•
I
mmediate Start:
Allows you to select to have the logs started shortly after the device is reconfigured.
•
Date/Ti
me:
Allows configuring the date/time at which the logs are to be started.
•
By Keyboard:
Allows you to enable a screen in the displ
ay where you can change the status of registrations to
enabled
. Therefore, the logs will
be started at that t
ime.
•
By Digital Input:
Allows you to configure it so that the
logs start from the digital input. This parameter is only available if the digital input is
enabled and confi
gured for Logs Control, so that the logs are started the moment the digital input is triggered. The log edge shall be configured
in the digital input parameters.
•
By Soft
ware:
Pro
vides permission for the logs to be started by a command from
NXperience Mobile
or
NXperience
.
•
Daily:
Allows a s
chedule to be configured so that logs are started every day at the configured time. A "Daily" log start mode also requires a
"Daily" log stop m
ode.
STOP MODE
•
Me
mory Mode Only:
Allows the configuration chosen for the Memory Mode to dictate the logs behavior: if Full Memory is chosen, the logs will
end as soon as th
ere is no more memory space; if Circular Memory is chosen, the logs will continue indefinitely.
•
Date/Ti
me:
Allow
s the user to configure the date/time at which the logs are to be stopped.
•
By Keyboard:
Allows you to enable a screen in the displ
ay where you can change the status of registrations to disabled. Therefore, the logs will
be stopped at that time.
•
By Digital Input:
Allows you to configure so that the logs are stopped from the digital input. This parameter is only available if the digital input is
enabled and conf
igured for "Logs Control", so that the logs are stopped the moment the digital input is triggered. The log edge shall be
configured in the
digital input parameters.
•
By Soft
ware:
Pro
vides permission for the logs to be stopped by a command from
NXperience Mobile
or
NXperience
.
•
Daily:
Allows a schedule to be configured so that logs
are stopped every day at the configured time. A "Daily" log stop mode also requires a
"Daily" log start m
ode.
The combinations all
owed for starting and ending logs are as follows:
Start Mode
I
mmediate
Date/Time
Digital Input
Keyboard
Software
Daily
Stop Mode
Memory Mode or Date/Time or by Keyboard or Software or Digital Input
Memory or Date/Time or by Keyboard or Software or Digital Input
Memory or Digital Input
Memory or Keyboard
Memory or Software
Daily
Table 06
– Stop modes of the logs
28/65