17.
TECHNICAL SPECIFICATIONS
CHARACTERISTICS
Input Channels
LOGBOX 3G
2 Analog
1 Digital
Co
mpatible Analog Signals
Internal Measurements
Analog Channel Input
I
mpedance
Pt100
Thermocoupl
e J, K, T, N, E, R, S and B, Pt100, 0-50 mV, 0-5 V, 0-10 V, 0-20 mA, 4-20 mA
Internal Temperature (NTC)
Battery Voltage
External Power Supply Voltage
Thermocouples / Pt100 / mV: > 2 M
Ω
mA: 15
Ω + 1.5 V
V: 1 M
Ω
Maximum compensation cable resistance: 25
Ω
Excitation current: 166
µA
Curve used:
α
= 0.00385
Logical levels
Maximum Voltage
Digital Input
Input Impedance
Input Current @ 30 VDC (typical)
Maximum Frequency (square wave)
Minimu
m Pulse Duration
Logical Level
"0": from 0 to 0.5 VDC
Logical level
"1": from 3 to 30 VDC
30 Vdc
270 k
Ω
150
µA
Dry Contact: 10 Hz, PNP: 2 kHz, NPN: 2 kHz
Dry Contact: 50 ms, PNP: 250
µs, NPN: 250 µs
Digital
Output
Displa
y
Resolut
ion
Me
mory Capacity
Logging Interval
Log Type
Log Trigger
Alar
ms
Internal Buzzer
Co
mmunication Interfaces
Soft
ware
GPS
1 NPN-type output
Maximum current that can be switched at output: 700 mA
3 lines, 4½ digits
Analog Signals: 15 bits (32768 levels)
Digital Signal: 16 bits (65536 levels)
140,000 logs (total)
1 second to 12 hours (recommended: 300 seconds)
Instant or Medium
Date/time, keypad, digital input, software command or SMS
10 alarms available (they may trigger the digital output, internal buzzer and send SMS)
Yes, it can be used for alarms
USB
Cellular Module (2G, 3G, SMS)
NXperience
(for desktop and notebook computers - locally via USB or remotely via
NOVUS Cloud
)
Available for models with GPS
Voltage: 10 VDC to 30 VDC
Po
wer supply
Po
wer Supply
Backup Batteries
Maximum Consumption: 700 mA
Typical Consumption: 70 mA
1 or 4 internal batteries depending on mod
el
*
Esti
mated Backup Battery Life
Operating Temperature
Enclosure
Protection Rating
Di
mensions
Typically 2 hours for the model with 1 battery
Typically 8 hours for the model with 4 batteries
Using Power Supply: -20 to 60 °C
**
Using Backup Power: 0 to 45 °C
***
ABS+PC
IP40
120 x 100 x 40 mm (without antenna)
Table 33
– Technical Specifications
* Risk of explosion: Internal batteries may only be replac
ed by the manufacturer or authorized service provider.
** Risk of explosion: Be careful with the device’s operating temperature. Extremely high or low temperatures can cause ruptures
and
leaks in backup batt
eries and cause damage.
*** Backup batteries will be charged only while the device is operating in the temperature range from 0 to 45 °C.
62/66