13.1.6.2
GROUPS
Fig. 19
– Add Group
•
Groups:
Allows f
or adding up to 5 groups with 20 contacts that can later be selected receive SMS messages related to the device alarm
situations.
•
Add
•
Delete
:
Allows for adding a group to the group list.
:
Allows for deleting a selected group from the group list.
•
Delete All
:
Allows for deleting entire group list.
13.1.7 ALARM CONFIGURATION
Fig. 20
– Alarms Screen
Alarm configuration allows for configuring up to 10 specific alarm conditions for the device to take certain actions. Among these actions, SMS
sending and buzzer and/or digital output triggering can be configured for when the configured alarm situation is reached.
•
List of Alarms:
Allows for selecting the desired alarm to be configured.
•
Tag:
Allows for configuring a name to be used as identification for the alarm. The field accepts up to 16 characters.
•
Operation:
Allows for comparing one of the variables measured by
LogBox 3G
against a value so that the device identifies the situation and
takes the actions configured for that alarm.
•
Hysteresis:
Allows for configuring a hysteresis for the configured alarm condition.
•
Alarm Actions:
o
Send SMS to
: Allows for linking groups and/or contacts to receive SMS messages when the device identifies the configured alarm situation.
o
Turn on Buzzer:
If selected, it configures the device to trigger the buzzer when it identifies the configured alarm situation.
o
Turn on Digital Output:
If selected, it configures the device to trigger the digital output when it identifies the configured alarm situation.
40/66