Recommended Action
Type / Cause
Error Message
Error Code
Verify the communication
environment and operating
status between the system
and the workstation.
If the error cannot be
removed, then press
(Power button) to turn the
power off, and then contact
the FSE..
Error 3. The recovery failed during
an attempt to resend the
transmission data from the
system to the workstation.
The system could not
resend the data
because the buffer
memory of the detector
overflowed during
analysis.
0x9344
Verify the communication
environment and operating
status between the system
and the workstation.
If the error cannot be
removed, then press
(Power button) to turn the
power off, and then contact
the FSE..
Warning. Communication
between the system and the
workstation was disconnected.
Communication
between the
instrument and the
workstation has timed
out.
0x9352
Avoid duplicate IP addresses.
If the error cannot be
removed, then press
(Power button) to turn the
power off, and then contact
the FSE..
Error 3. The IP address of the
system overlaps with that of
another system.
The IP address of the
instrument is
duplicated.
0x9356
Clear the error. There will be
no alarm sound or error
message. Work while taking
sufficient care for avoiding
burns.
Type: Warning. The column oven
door is open with the column
oven temperature at 50 °C min.
The column oven door
is open when the
column oven is at high
temperature. Close the
column oven door as
soon as possible in
mind, such as burns.
0x9400
ExionLC
™
100 System
Hardware User Guide
209 / 219
RUO-IDV-05-1839-C
Error Messages