Recommended Action
Type / Cause
Error Message
Error Code
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
Could not detect the
rotation direction
home position of the
autosampler measuring
pump.
0x9203
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
Temperature control is
turned off because the
autosampler rack L
heater has overheated
and the thermal fuse
has blown.
0x9208
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
Temperature control is
turned off because the
autosampler rack R
heater has overheated
and the thermal fuse
has blown.
0x920A
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
The leak sensor in the
autosampler low
pressure valve has
detected an abnormal
condition.
0x920B
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
The leak sensor in the
autosampler low
pressure valve is
malfunctioning.
0x920C
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
The leak sensor in the
autosampler high
pressure valve has
detected an abnormal
condition.
0x920D
Press the
Power OFF
button in the error window
to turn the power to the
system off, and then contact
the FSE.
Fatal error
The leak sensor in the
autosampler high
pressure valve is
malfunctioning.
0x920E
ExionLC
™
100 System
Hardware User Guide
203 / 219
RUO-IDV-05-1839-C
Error Messages